3.1.1 - Low coolant level
This image appears when the coolant level in the tank is low. The treatment process will stop
otherwise could damage the peltiers inside the applicators.
To solve the problem, proceed with filling the cooling system with antifreeze of 10-20%, follow the
process explained in installation and maintenance section.
In case of filling the equipment and the error persists, check the operation of the tank level sensor.
With full tank, measure continuity on the wire marked 0 (tank level sensor) and check that the dry
contact is closed.
Level sensor connector
If the contact does not close, try to measure again while moving the floater state downwards. If in this
case the contact closes, change floater position upside down. On the contrary, if it does not close,
proceed with replacement of the deposit tank.
In case of correct operation of the level sensor, proceed to replace Main CPU.