CleanStation Dimension SRS-DT3
TM
Jet action cleaning
Troubleshooting
If CleanStation
TM
does not run - no power to the display.
•
Verify the power switch is turned on, the system is plugged into a
power source, and that power is available.
• Verify adequate water level in the tank. If water level in the tank
drops below 1 inch (approximately) from the top of the tank,
CleanStation
TM
is designed to power itself off.
• Check for a blown fuse. There is one fuse located behind the tank,
facing down. After having thoroughly drained and unplugged the
system from it's power source, tip the unit so that it's resting on it's left
side (when viewed from the front panel). Locate the fuse (see Figure
5) looking up from the bottom of the system, behind the tank. After
completely removing the fuse, check with an ohm meter. A blown
fuse usually indicates some other problem … with the system still on
it's side, spot check that all wires are undamaged.
Figure 5
main fuse
pump circuit
breaker
If CleanStation
TM
heats with no water agitation.
• Check for tripped pump circuit breaker. There is a circuit breaker located behind the tank, facing down. After having thoroughly
drained and unplugged the system from it's power source, tip the unit so that it's resting on it's left side (when viewed from the front
panel). Locate the circuit breaker (see Figure 5) looking up from the bottom of the system, behind the tank. The circuit breaker has
tripped if the position of the breaker switch is showing white. Reset the breaker by pushing it in.
• If the pump motor feels hot (to the touch), the temperature sensitive thermo switch may be tripping due to high heat. Remove any
obstructions from the drain filter and jets (refer to the maintenance section) and allow the pump to cool down before restarting.