6- 1
6. Parameter Settings
6-1 Setting the SET No. Parameters
Parameter name
Setting values
Remarks
Basic settings (BASIC)
HIGH setting value (HI.SET)
Sets a HIGH setting value for pass/fail judgment.
−199.9999 to 199.9999 (mm)
[Default: 5.0000]
LOW setting value (LO.SET)
Sets a LOW setting value for pass/fail judgment.
−199.9999 to 199.9999 (mm)
[Default: 1.0000]
Hysteresis (HYSTER)
Judgment may be unstable due to measurements
fluctuating around the HIGH or LOW setting value. A large
hysteresis value can prevent this chattering effect.
0.0000 to 199.9999 (mm)
[Default: 0.0030]
Teaching mode (TEACH)
Sets the method for automatic teaching of tolerance.
TCH.1 (One-point) / TCH.2 (Two-point) / TCH.3
(Three-point)
[Default: TCH.1]
Tolerance <±> (TOL <±>)
When executing the one-point teaching, this parameter is
set in order to provide a tolerance from which the HIGH
and LOW setting values are derived, with the master work
measurement taken as the center value.
0.0000 to 199.9999 (mm)
[Default: 0.1000]
Preset value (PR.VAL)
Sets the reference value when the preset function is turned
ON.
−199.9999 to 199.9999 (mm)
[Default: 0.0000]
Select preset data (PR.OBJ)
When the preset function is turned ON, this parameter
selects the type of current value (NORM.V or JUDGE.V) to
which the offset value is applied. NORM.V (normal
measurement value) / JUDGE.V (judgment value)
[Default: NORM.V]
Preset save (PR.SAVE)
If Preset save is set to ON, the preset and offset values are
written to an EEPROM so that they are retained after the
power is turned OFF.
ON / OFF
[Default: OFF]
Measurement direction
(DIRECT)
Selects between normal and reversed display of the
measurement value when the spindle of the sensor head is
pushed in during measurement.
NORMAL (Normal) / REVERS (Reverse)
[Default: NORMAL]
Lever ratio (LEVER)
If a measurement point is on a lever, the measurement
value can be displayed as detected value
×
lever ratio.
0.1 to 100.0
[Default: 1.0]
Response time (SPEED)
Sets the period between the time the sensor head starts
measurement and that at which the measurement value is
determined.
3 / 5 / 10 / 100 / 500 / 1,000 (ms)
[Default: 100]
Output behavior (OUTPUT)
Selects behavior of judgment output.
N.O. (Normally open) / N.C. (Normally closed)
[Default: N.O.]