6-1-3 Hysteresis (HYSTER)
Parameter name
Setting value
Remarks
Hysteresis (HYSTER)
Judgment may be unstable due to measurements fluctuating
around the HIGH or LOW setting value. A large hysteresis value
can prevent this chattering effect.
0.0000 to 199.9999 (mm)
[Default: 0.0030]
With HYSTER set to 0:
Chattering makes judgment output unstable.
With HYSTER set to non 0:
Judgment output becomes stable.
Operation procedure
Note
HIGH and LOW share the same hysteresis value.
Set hysteresis to a value smaller than the difference between the HIGH and LOW setting values. If
hysteresis is set to a value equal to or exceeding the difference between the HIGH and LOW
setting values, the GO judgment will not be output.
■
■
HIGH設定値
LOW設定値
HIGH
GO
LOW
HIGH setting
value
LOW setting
value
HIGH設定値
LOW設定値
HIGH
GO
LOW
HIGH setting
value
LOW setting
value
2 秒長押し
/
選択
キャンセル
桁移動
数値変更
確定
HI. SE T
LO . S E T
T E ACH
T O L <±>
PR. VAL
ADVANC
CALC
CO PY
BANK
CALIB
PR. O BJ
PR. S AVE
DIRE CT
LE VE R
S PE E D
O UT PUT
MAINT E
RE S E T
Long press
for 2 s
Select
Commit
Cancel
Shift between
digits
Change numerical
values