6- 15
6-1-11 Response Time (SPEED)
Parameter name
Setting value
Remarks
Response time (SPEED)
Sets the period between the time the sensor head starts
measurement and that at which the measurement value is
determined.
This parameter is effective for stabilizing measurement values.
3 / 5 / 10 / 100 / 500 / 1,000 (ms)
[Default: 100]
Response time (ms)
No. of cycles per average
Averaging time (ms)
Internal value update
period (ms)
3
1
1
1
5
3
3
1
10
8
8
1
100
96
96
4
500
480
480
16
1,000
960
960
32
Operation procedure
Note
When the reset input is turned ON, the display shows an indefinite value "------" until the device
reaches the number of cycles per average.
■
■
When response time is set to 10 ms:
Data
n-7
Data
n-6
Data
n-5
Data
n-4
Data
n-3
Data
n-2
Data
n-1
Data
n
Data
n+1
Data
n+2
Data
n+3
Sampling time (1 ms)
Average over eight cycles
→
internal value
Average over eight cycles
→
internal value
Averaging period (8 ms)
Response time (10 ms)
Internal value update
period (1 ms)
2秒長押し
/
選択
キャンセル
選択
確定
HI. S E T
LO . S E T
HYS TE R
T E ACH
T O L <±>
ADVANC
CALC
CO PY
BANK
PR. VAL
PR. O BJ
PR. S AVE
DIRE CT
LE VE R
O UT PUT
MAINT E
CALIB
RE S E T
Long press
for 2 s
Select
Commit
Cancel
Select