6-2-6-2 Scaling Upper Limit/Scaling Lower Limit
Parameter name
Setting value
Remarks
Scaling upper limit (ANA.HI)
This can be set only when analog scaling is set to FREE.
It sets the judgment value that outputs a current of 20 mA
−199.9999 to 199.9999 (mm)
[Default: 10.0000]
Scaling lower limit (ANA.LO)
This can be set only when analog scaling is set to FREE.
It sets the judgment value that outputs a current of 4 mA.
−199.9999 to 199.9999 (mm)
[Default: 0.0000]