Remote Control Reference
4-11
:SEQuence:STEP:CURRent? {<step>}
:SEQuence:STEP:DWELl {<step>},{<time in ms>|MIN|MAX}
:SEQuence:STEP:DWELl? {<step>}
:SEQuence:STEP:RAMP {<step>},{<time in ms>|MIN|MAX}
:SEQuence:STEP:RAMP? {<step>}
:SEQuence:STEP:VOLTage {<step>},{<voltage>|DEF|MIN|MAX}
:SEQuence:STEP:VOLTage? {<step>}
:SEQuence:STEP? {<step>}
OUTPut:SEQuence[:STATe] {0|1|OFF|ON}
Enable/Disable output sequence function.
OUTPut:SEQuence[:STATe]?
Return 0 if output sequence is disabled, and return 1 if output sequence is enabled.
OUTPut:SEQuence:MODE {0|1|2}
Set mode of output sequence. 0 is voltage only. 1 is current only. 2 is both voltage and current
are controlled by the output sequence.
OUTPut:SEQuence:MODE?
Query sequence mode. Return 0, 1 or 2.
OUTPut:SEQuence:CYCLe {<cycle>}
Set output sequence repeat cycle. The value of cycle is 0~65535. 0 means always repeat.
OUTPut:SEQuence:CYCLe?
Query output sequence repeat cycle.
OUTPut:SEQuence:SETup {<start step>},{<stop step>}
Set start step and stop step of output sequence. Step is S0~S7.
OUTPut:SEQuence:SETup?
Query start step and stop step.
OUTPut:SEQuence:STEP:VOLTage
{<step>},{<voltage>|DEF|MIN|MAX}
Set voltage of a step. Step is S0~S7.
OUTPut:SEQuence:STEP:VOLTage? {<step>}
Query voltage setup of a step. Step is S0~S7. Return voltage in Volts.
OUTPut:SEQuence:STEP:CURRent {<step>},{<current>|DEF|MIN|MAX}
Set current of a step. Step is S0~S7.
OUTPut:SEQuence:STEP:CURRent? {<step>}
Query current setup of a step. Step is S0~S7. Return current in amps.
OUTPut:SEQuence:STEP:RAMP {<step>},{<time in ms>|MIN|MAX}
Set ramp time of a step. Step is S0~S7. The ramp time allows you to set from 0 to
3599999ms.
OUTPut:SEQuence:STEP:RAMP? {<step>}
Query ramp time of a step. Step is S0~S7. Return ramp time in ms.
OUTPut:SEQuence:STEP:DWELI {<step>},{<time in ms>|MIN|MAX}