Programmable DC Power Supply 62000L Series User’s Manual
3-8
62000L have to be wired to your DUT’s + and - terminals. Third, you need to enable the sense
option via the front panel.
1. Wire the 62000L’s output to your DUT:
You can connect the & – of the power supply, and DUT’s + & - through either
banana leads, or bare leads.
Note
When you use bare leads to connect the power supply, please rotate the + & -
terminals of the power supply counter-clockwise to release some spaces for bare
leads’ insertion. Then rotate the terminals clockwise to fasten the connection.
Connection via Banana Leads Connection via Bare Leads
2. Wire the 62000L’s sense to your DUT:
The terminals of the positive sense (S+) and the negative sense (S-) are located with the
terminals of the positive transfer (T+), the negative transfer (T-), the positive recipient (R+),
and the negative recipient (R-). The terminals (T+/T-/R+/R-) are used for Master-Slave
control.
To wire the sense terminal of your power supply, you need to use a screwdriver with a flat
head whose diameter is equivalent to or less than 3 millimeters. With the screwdriver
pressing on the red area, you can insert a lead into the terminal as shown below.
Negative Sense Connection Positive Sense connection
3. Enable the sense via the front panel operation:
To enable the remote sense, you have to turn off the output first and press the Menu
button once. The SENSE symbol will blink. Press the Menu button once again. The
display will show a blinking INT. Rotate the knob to display the blinking EXT symbol.
Press the Menu button to finish the setting. The output will return to the meter mode. The