9-3
9-3 Shift adjustment of measured values
This function is provided to adjust the shift (parallel move) quantity of data display and digital recording.
This function is used for correcting the dispersion of sensors, an input converter, etc.
This adjustment can be done every channel in two ways.
1. Set a correction value by channel parameter setting [CH PARAM.].
For details, see 8-3.
Display the above setting parameter entry screen, and input a correction value to the [Correction] parameter
input area by numeric keys. Bring the cursor to Set and terminate this entry by pressing Enter key.
2. Set a correction value by measured value zero-span adjustment [INPUT ADJ.]
For details, see 9-2.
Display the setting entry screen in the left figure, input [1]
to [a] parameter input area, and input a correction value
to [b] parameter input area by numeric keys.
Bring the cursor to Set and terminate the entry by
pressing Enter key.
(Caution 1)
If a correction value is set in channel parameter setting, a correction value is added when the correction
value is set again in zero-span adjustment of measured value.
(Caution 2)
Be careful since the value shifts by an adjusted value during calibration.
(Caution 3)
Feed power to the unit for longer than 30 minutes at least for the purpose of stabilizing the unit before
starting the adjustment.
(Caution 4)
A range setting-free channel is not selectable.