12-2
Alarm printing
(when option is added)
:
Generation or cancellation mark in tag (Channel number in case of non-setting), kinds of alarms,
generation and cancellation time, level, and analog recording
Scale printing
:
Minimum value, maximum value, tag (Channel number in case of non-setting)
List printing
:
All setting parameters printing, designated parameters printing
Message printing
:
Optional setting within 80 characters Time is printed
Channel designation is also possible (within 75 characters).
Data printing
:
Latest data are printed out by Data P key.
Differential operation
(
∆
T)
:
A difference between optional channels or a difference from an optional set value
Display section
:
Dot matrix color LCD panel (6.2 inches) of 640 x 240 dots
LED status: Alarm (ALM) red, Memory card (PC CARD) green, Recording ON/OFF (REC ON) green
Display contents
:
Selectable out of one-point fixed or sequential display, all points collective display, 12-point fixed or
sequential display (Unit number display: Type code, serial number, software version number)
Status
・・・・・・・・
Key lock, chart end, setting decision (format error or setting error), etc.
LED status
・・・・
Lights in the occurrence of an alarm, memory card operation, and recording ON
Skip function
:
Skip channel is optionally settable (Settable independently of recording skip)
Operation assembly
:
Operation
keys
・・・
Entry and setting function keys of various setting parameters by conversational
system, recording control keys, alphabetic keys, ten keys, etc.
Card setting function
:
Memory card (option) is used.
Exclusive engineering
port
:
USB port ( B type) which is set to PC to set and control all setting parameters.
For this operation, exclusive engineering software (option) for PC is necessary.
Alarm system (option)
:
Individual setting, higher limit, lower limit
Setting system
:
Individual setting by key operation at each point, 4 levels/channel
Alarm display
:
Red LED status lamp lights in the occurrence of an alarm.
Alarm printing
:
Tag (channel number in case of non-setting), kinds of alarms, level, generation and cancellation time
at the right end Generation and cancellation marks on analog recording
Alarm output
:
3 kinds of 12 points, 24 points, and 36 points (AND output and OR output can be designated)
Contact capacity: 100V AC 0.5A resistive load, 240V AC 0.2A resistive load, 30V DC 0.3A
resistive load
Power supply
:
100V
AC
-
240V AC (50/60Hz)
Power consumption
:
Approx.
100VA
Working temperature range
:
0 to 40
℃
*
during transportation and storage : -20 to 60
℃
Working humidity range
:
20 to 80%RH *during transportation and storage : 5 to 95%RH (
no condensation)
Countermeasures against power
interruption
:
Set values protection
・・・
EEP, ROM
Clock
・・・
Backed up for longer than 5 years by lithium battery
Insulation resistance
:
500V DC, 20M
Ω
or higher between primary terminals and protective conductor terminals
500V DC, 20M
Ω
or higher between secondary terminals and protective conductor terminals
500V DC, 20M
Ω
or higher between primary terminals and secondary terminals
Dielectric strength
:
1500V AC, 1min between primary terminals and protective conductor terminals
500V AC, 1min between secondary terminals and protective conductor terminals
1500V AC, 1min between primary terminals and secondary terminals
(Note 1) Primary terminals: Power terminals, alarm output terminals, output relay terminals
Secondary terminals: Measuring input terminals, communication terminals, external drive
terminals
(Note 2) Short all primary and secondary terminals before starting the insulation resistance and/or
dielectric strength test, otherwise the unit may be broken
Internal illumination
:
White
LED
System error alarm
:
[Fail] display and relay output (Relay output (option) contact capacity: 100V AC 0.5A, 200V AC 0.2A)
when the unit is in trouble
Chart end detection
:
Chart End] display and relay output at the end of chart
(Relay output: option Contact capacity: 100V AC 0.5A, 200V AC 0.2A)
Key lock
:
By operation key operation (A password input is necessary) [Key Lock] display
Case
:
Front
door
・・・
Aluminum die cast (Operation assembly is made of ABS resin)
Rear case
・・・
Normal steel plate
Painting color
:
Door and case
・・・
Color equivalent to DIC546 1/2
Operation assembly
・・・
Color equivalent to DIC549
Mounting method
:
Panel embedded mounting, Position
・・・
Horizontal in lateral direction
Forward tilting 0 degree, Backward tilting
・・・
Within 15 degrees
Terminals
:
Input terminals, alarm output terminals
・・・
M3.5, Others
・・・
M4
Weight
:
Approx.
15kg
Содержание LE5100
Страница 2: ......
Страница 89: ......
Страница 90: ...32 8 KUMANO CHO ITABASHI KU TOKYO 173 8632 Telephone 81 3 3956 2171 Facsimile 81 3 3956 0915 Printed in Japan ...