8-32
8-17-6 Setting of measuring data saving by PC card [DATA WR.]
(
Note 1
)
If the start trigger is set to be [None], measuring data is
not saved even if the start time is set.
Set the start trigger to a [Time] without fail when setting
the start time. If a set start time is earlier than the
present time and minutes, data are started saving at
the set start time in the next day.
(
Note 2
)
Set the start trigger to the position other than [Note] if
start triggering is done by keying operation (Shift +
Rec. keys).
Even if the start trigger is set to a [Time], the start by
keying operation (Shift + Rec. keys) takes preference
of the start by the designated time.
This also applies to the end trigger, correspondingly.
Neither start triggering nor end triggering is executed
by Shift + Rec. keys operation only.
Refer to
※
1)
on the right side.
Both start triggering and end triggering can be
executed irrespective of the display screen.
(
Note 3
)
If the PC card capacity does not conform to the
recording time even if a recording time is designated,
the saving work is completed to meet the capacity.
When parameter is loaded during saving measuring
data, the invalidity data is saved.
(
Note 4
)
When select [output relay] in start trigger or end
trigger, option of contact output (such as alarm
output) is needed. When select [external contact],
option of external drive is also needed.
①
Select DATA WR.
②
Press Enter key to select the parameter entry mode of an item to be set.
A window opens and the setting parameter entry screen is displayed.
(For the setting parameter entry screen, refer to
④
)
③
For stopping the setting, the screen returns to the last one, each time
Esc key is pressed once.
④
Move the cursor to the item to be set by
◄
・
►
・▲・▼
key.
The cursor moves in the sequence of Format
→
Start TRG
→
Start Hour
→
Minute
→
Relay No.
→
End TRG
→
Rec.time Hour
→
Minute
→
Interval
→
pre TRG
→
Set.
1. For parameter input area of the Start Hour and relay No. and Minute
and Rec.time Hour and Minute and parameter input area of pre TRG,
input setting parameters by numeric keys.
Select one output [800] to [836] for relay No. Pretrigger is [optional
setting 0] to [100] by every 1.
**Pretrigger Example: If pretrigger is set [10], start recording the [10]
previous data from the time when trigger is ON.
2. For the Format, and Start TRG, and End TRG, and Interval parameters
input area, input their setting parameters by numeric keys. A window
opens. Select them by using
▲・▼
key. For each selection, refer to
the following table.
3. After the end of setting on this screen, bring the cursor to
Set
.
⑤
Terminate the setting (change, addition) by pressing Enter key.
Press Esc key if the set contents are not saved.
・
Format
・
Start TRG
・
End TRG
Binary
None
Key (Shift + Rec.)
Text
Key (Shift + Rec.)
Time
Time
Relay
No.
Relay
No.
Ext.
Ext.
Binary : Extension is [l5f]. Data analysis software is necessary to
reproduce.
Text : Extension is [txt]. Possible to reproduce with Excel (Microsoft Co.)
etc. if necessary.
・
Interval
0.1s 0.2s 0.5s 1s 2s
3s 5s 10s
15s
20s
30s
1
m
2
m 3m 5m
10
m
15
m
20
m
30
m
60m
※
1
)
For executing the Start TRG and End TRG by keying operation;
Start TRG
:
The following message is displayed by pressing Shift +
Rec. keys.
[Do you start recording to the memory card?]
Press [Enter] key in case of yes or press [Esc] key in case of no.
When starting the record, display [R] left next to the residual quantity.
End TRG
:
The following message is displayed by pressing Shift + Rec.
keys.
[Do you finish recording to the memory card?]
Press [Enter] key in case of yes or press [Esc] key in case of no.
When finishing the record, disappear [R] left next to the residual quantity.
Start
Set?
①
Esc key
NO
③
End
YES
Setting (change, addition)
by using
◄
•
►
•
▲
•
▼
keys, numeric keys, etc.
④
②
Enter key
Setting ends
⑤
DATA WR.
Enter ke
y