34
-
5. LOADING
5.3 Chart Loading
(1) Chart Cassette Removal
1) Open the Door.
2) Remove the Chart Cassette.
Grasp the handle on the chart cassette and pull it out.
(2) Chart Paper Setting
1) Open the Chart Paper Holder.
Open the front chart holding axis in the front side and the rear
stripper plate in the rear side.
2) Prepare a Chart Paper.
To prevent double feeding, disentangle the both ends of paper
well.
Door
Handle
Guide
Rear stripper plate
Front chart holder axis
[For dot-printing type]
Pull out the chart cassette while RECORD switch is off
(recording off) or power is off for the dot-printing type.
If the chart cassette is pulled out in the state of the dotter
down during power off (The state which top of the dotter
touches the chart. This state may happen if the power is
turned off while the dotter operates.), the dotter may be
damaged.
When the dotter is down, turn the power on once, and when
“initial operation of the dotter” ends, turn the power off (Refer
to 6.1 Power On/Off.).
Check the State of Dotter
[For pen type]
Pen lift is automatically operates itself at the time of pulling
out the chart cassette and putting back into the inner unit
Pen lift lever operation is unnecessary, however, there might
be a possibility that ink stain of the cartridge pen left on the
chart. Let the cartridge pen UP status by the pen lift lever
then load the chart prevent leaving ink stain on the chart.
Pen Lift Lever Operation