18
-
3. FRONT
3.2 Display Operation Part
[Pen type]
(1)
Detail of Display Operation Part
(2)
Function of Display Operation Part
1. Switch
{
1
1st pen INDICATE: Switching of ON/OFF of Indication/Recording of
the 1st pen
{
2
2nd pen INDICATE: Switching of ON/OFF of Indication/Recording of
the 2nd pen
{
3
3rd pen INDICATE: Switching of ON/OFF of Indication/Recording of
the 3rd pen
○
4 Refer to 6.2 Operation/Setting Change (3) DIP Switch Function
2. Setting/operation key
{
A
FEED:
Feeding of chart papers with key operation
{
B
MODE:
Operation on transition to Setting mode
{
C
Left ( ):
Switching/selection operation on various
operation/setting
{
D
Right ( ):
Switching/selection operation on various
operation/setting
{
E
ENTRY:
Fixing operation for selection parameters on
various setting
3. Indication lamp
{
ⅰ
Power supply:
Indication of power supply On/Off
{
ⅱ
Alarm:
Indication of alarm occurrence (for all channels)
{
ⅲ
Chart speed:
Indication of chart speed and unit
{
iv
Setting mode:
Indication of various setting modes
{
v
Setting parameter: Indication of selection parameters on setting mode
(Chart speed lamp is used also for this purpose.)
○
vi Pen selection: Indication of selection pen number on setting mode.
(*) Engineering connector is for maintenance. It is not used on normal operation.
(*2) it is a screw hole for front cover fixing. It is not used on
normal specification (refer to 6.9 Operation Setting Function
Depending on Application (2) Key Lock Function
Another Key Lock Method).
Setting information and various statuses information of
the instrument is stored in the nonvolatile memory
(EEP-ROM) within the instrument.
When the setting information of the instrument is
changed, the alarm lamp blinks for about 3 seconds.
When the status information of the instrument is changed,
the alarm lamp blinks for about 1.5 seconds.
Blinking of Alarm Lamp
Avoid closing the door in the state of front cover opened.
If the door is closed in the state of the front cover opened,
the mechanism of the front cover allows the cover to be
lifted to the direction for closing to prevent damage;
however, behavior for protection is not guaranteed. If the
door is closed forcedly or fast, it may be damaged.
Front Cover Protection
1) State of front cover closed
2) State of front cover opened
{
ⅰ
Power supply
{
ⅱ
Alarm
{
ⅲ
Chart speed
{
A
FEED
{
3
3rd pen INDICATE
{
1
1st pen INDICATE
{
2
2nd pen INDICATE
(*2)
○
viPen selection
(*1) Engineering connector
○
4 DIP switch
{
C
Left
{
D
Right
{
E
ENTRY
{
B
MODE
{
iv
Setting mode
{
v
Setting parameter (only during Setting mode)