Device
Mod. Clenny
®
A Kid
Power supply:
12V DC 1.25A
Max pressure:
2 ± 0.3 bar
Compressor air delivery:
approx 9 l/min
Noise (at 1 m):
approx 51 dB (A)
Operation:
Use 20 minutes ON/ 20 minutes OFF
Operating Conditions:
Temperature:
min 10°C; max 40°C
Air humidity:
min 10%; max 95%
Atmospheric pressure:
min 69KPa; max 106KPa
Storage conditions:
Temperature:
min -25°C; max 70°C
Air humidity:
min 10%; max 95%
Atmospheric pressure:
min 69KPa; max 106KPa
Compliant with Directive 93/42 EEC
Dimensions:
14.7 (W) x 14.7 (D) x 7 (H) cm
Weight:
0.58 Kg
Power supply
Only use power supply mod. UME318-1215 (cod. 17111)
Primary:
100/240V~ 50/60Hz
Secondary:
12V DC 1.5 A
APPLIED PARTS
Type BF applied parts are:
patient accessories (C2, C8)
Clenny A 2ME nebuliser cup "2 delivery methods"
Minimum medication capacity:
2 ml
Maximum medication capacity:
8 ml
Operating pressure
(1)
(with neb.):
0.7 bar
TECHNICAL SPECIFICATIONS
Delivery mode selector C6.4
Mode 1
Mode 2
(press Max)
(1)
Delivery:
approx. 0.18 ml/min
approx. 0.42 ml/min
(2)
MMAD:
2.44 μm
2.95 μm
(2)
Breathable fraction
<
5 μm
(FPF):
80.7%
74.7%
(1) Data based on using 0.9% NaCl saline solution, according to internal procedure.
(2) In vitro characterisation performed with NGI instrument certified by TÜV Rheinland LGA Products GmbH - Germany
in compliance with the European Standard for Aerosol therapy apparatuses EN 13544-1. Further details are available
on request.
PROBLEM
CAUSE
SOLUTION
The device
does not work
The power cable has not been
correctly inserted into the sock-
et of the device or the power
socket
Correctly insert the power cable
in sockets.
The device
does not nebulize
or nebulizes
insufficiently.
The medication has not been in-
serted in the nebulizer
Pour the right amount of medi-
cation in the nebulizer
The nebulizer has not been
properly fitted.
Disassemble the nebulizer and
reassemble it correctly as shown
in the connection diagram on
the cover.
The nebulizer nozzle is clogged
Disassemble the nebulizer, re-
move the nozzle and perform
cleaning operations Failure to
remove medication deposits
from the nebulizer affects its ef-
ficiency and operation.
Strictly comply with the instruc-
tions contained in the CLEAN-
ING, SANITISATION AND DISIN-
FECTION chapter
The air pipe is not correctly con-
nected to the device
Ensure the device’s accessories
are properly connected to the air
intake (see connection diagram
on the cover).
The air pipe is bent, damaged or
twisted
Unwind the pipe and ensure it is
not crushed or punctured. If nec-
essary replace it.
The air filter is dirty
Replace the filter
The apparatus is
noisier than usual
The Filter is not correctly
inserted in its housing
Fully and properly insert the Fil-
ter in its housing
Содержание Clenny A
Страница 51: ...49 v A A1 A2 A3 A4 5 A6 B C C2 1 6 Clenny A 2ME 2 C6 1 C6 2 C6 3 C6 4 C7 1 12 C8 1 12 D 1 6...
Страница 52: ...50 1 L A5 2 C6 1 3 C6 3 C6 1 4 400 93 42 CE...
Страница 53: ...51 1 SoftTouch CLENNY A 2ME 2 C6 4 MAX C6 4 MAX Clenny A 2ME MAX 1 1 12 5 A1 6 7 1 12...
Страница 55: ...53 A3 II EN 10993 1 93 42 EEC CE 1907 2006 CE 93 42 CEE IP21 12 BF ON OFF...
Страница 57: ...55 k...
Страница 58: ...56 EN 60601 1 2 2015 2012 19 CE 50 22 1997...
Страница 59: ...57 A A1 A2 A3 A4 A5 A6 B C C2 Clenny A 2ME C6 2 C6 1 C6 2 C6 3 C6 4 C7 12 1 12 1 C8 D 6...
Страница 60: ...58 A5 L 1 C6 1 2 C6 1 C6 3 3 4 400 42 93...
Страница 61: ...59 12 1 Soft Touch CLENNY A 2ME 2 C6 4 MAX C6 4 MAX Clenny A 2ME MAX A1 5 6 B 7 12 1...
Страница 64: ...62 A3 II EN10993 1 EEC 93 42 CE 1907 2006 CE CEE 93 42 IP21 12 FB ON OFF...
Страница 65: ...63...
Страница 66: ...64 CE 19 2012 CE 19 2012 EN 60 60601 1 2 2015 EMC RE...
Страница 67: ...65 5 MEMO MEMORANDUM NOTIZEN NOTAS...