5
1.1.3.
Precautions for use
The operator and/or the responsible authority must
carefully read and clearly understand the various
precautions to be taken in use.
Do not use the instrument in an explosive
atmosphere or in the presence of inflammable gas or
smoke.
Do not use the instrument on networks with a rated
voltage or category higher than those mentioned.
Do not use the instrument if it seems damaged,
incomplete or incorrectly closed
Keep your hands and fingers away from the unused
terminals of the device. When handling sensors or
test probes, do not place fingers beyond physical
finger guard.
All elements on which the insulation is damaged
(even partially) must be put out of service for repair
or disposed at waste.
Respect the environmental conditions of use.
Use
personal
protection
equipment
when
conditions require it.
When handling the leads, the probe tips, and the
crocodile clips, keep your fingers behind the
physical guards.
Comply with the environmental conditions of use.
2.
GETTING STARTED
Delivery condition
2.1.
Check completeness of the delivery against your order.
Delivered in a cardboard box with:
-
Operating directions in 11 languages on CD-ROM with SX-DMM software
-
Getting started guide on paper (available on CD)
-
1 set of safety leads (red and black) with double insulation probe tip (
of the probe tips: 4mm) 1,000V CAT III 20A
-
1 set of 4 AA/R6 Ni-MH storage batteries
-
1 5VDC, 2A mains/USB adapter (100-240V, 50/60Hz, 0.5A) with a USB power cord
-
1 statement of manufacturer's measurements
-
Optical USB communication lead
-
1 carrying bag
Accessories and spares
2.2.
-
Current clamps (see CHAUVIN ARNOUX catalogue)
-
Two-wire Pt100 temperature probe (HX0091)
-
Two-wire Pt1000 temperature probe (HA1263)
-
K thermocouple with banana adapter (P011021067)
-
Metrology software for Windows (HX0059B)
-
Set of rechargeable batteries (HX0051B)
-
HV probe (SHT40KV)
-
CMS clamp (HX0064)
-
Bluetooth key (P011102112)
-
Multifix adapter for DMM (P01102100Z)
-
External charger for Ni-MH storage batteries
-
Fuse 1,000V 11A >20kA 10x38mm (Get in touch with our Manumesure Regional Technical Centre)
-
Kit of test accessories for DMM (P01295459Z)
-
Carrying case with Multifix (HX0052C)
For the accessories and spares, consult our web site:
First use
2.3.
2.3.1.
Replacement (primary or storage batteries)
1.
Disconnect the instrument from all sources of current
2.
Unscrew the 3 screws on the back
3.
Open the lower housing with a screwdriver
4.
Remove the seal protecting the batteries (primary or rechargeable)
5.
Insert the batteries, paying attention to the polarity
6.
Close the housing and screw the screws back in
7.
Check the type in Setup/Pw supply/type (alkaline or NiMH)