26
5.2.2.
Procedure
1.
Press function A, then select the coupling according to your measurements: AC,
DC, AC+DC (AC+DC is default)
2.
Connect the black lead to the COM terminal, the red lead to A, and the probe tips
in series between the source and the load as shown below:
3.
Read the measurement indicated on the main display unit. Look up the graph of
the trend values >1min28s if it is activated. Look up the secondary measurements if they
are activated (activated by Meas…
MEAS+).
4.
It is possible to display the specifications of the range for metrology or a RELative
measurement.
5.
Surveillance of voltage
SURV
or Recording
MEM
of data internal to the
multimeter
"OL" is displayed if the current available I >20A.
6.
View the waveform of a current in the 10 Hz
– 600 Hz band
Current measurement with clamp
5.3.
5.3.1.
Connecting the multimeter
Current clamp with current output connected to the multimeter
Current clamp with voltage output connected to the multimeter
5.3.2.
Main measurement
To avoid opening a circuit, we recommend measuring the current with a current clamp, output A or V (Ax
function).
The clamp function includes a precise ratio xxxx.XA/xxxx.XV or XA, making it possible to connect a broad range
of current clamps that you will find in the CHAUVIN ARNOUX catalogue; it is however necessary to verify that the
input/output range of the clamp matches the ranges available on the multimeter.