4.1.1.
SETUP 1/3: general configuration of the multimeter
:
utility for adjusting the lighting, the standby mode, the audible beep of the
keys, the language, and the internal clock on 2 configuration levels.
:
for communication and adjustments of the IR/BT type, then the
IR rate in baud, and the protocol, MODBUS or SCPI.
: identifies the type of internal power supply used in the instrument, either Ni-
MH battery (capacity) or alkaline battery.
Util 1/2: Display
selection of 3 levels of back-lighting of the display unit in order to limit the
power consumption of the multimeter, as follows: Eco, Normal, Max
The default level of extinction of the back-lighting is ECO, after 1min if there has been no action
on the front panel of the multimeter. An internal accelerometer makes it possible to wake up the
multimeter by simply touching the product with the adjustment selected.
validation (default: yes) or not of automatic switching off after 30 min, if there
has been no action on the front panel of the multimeter.
In the SURV, MEM, or Communication mode, automatic switching off is not validated.
For your safety, automatic switching off is disabled when the quantities measured (voltage,
current) on the input exceed the danger thresholds.
: validation (default) or not of the emission of an audible signal (beep) when :
-
a key is pressed
-
there is a voltage on the "V" input exceeding 60 VDC or 25 VAC,
-
a stable measurement is captured in AUTO HOLD
The audible signal is maintained even when the buzzer is deactivated:
-
In a continuity test
-
When a range is exceeded (voltage or current)
-
For a measurement of 10A or more
-
When there is an incompatibility between the connections of the leads and the function selected
-
When the supply voltage (battery) is too low blinking of the red batt indicator
-
The audible signal is maintened when the function is changed while recording is in progress (low pitched beep)