46
Max. open-circuit voltage
3,5 V max.
28 V
Indication of overshoot
in the reverse direction
in the reverse direction
“Active” protection by PTC thermistor
1414 Vpk
1414 Vpk
Audible continuity
8.9.
In this setting, you measure a resistance up to 1000
, with continuous 4kHz audible indication
Range
Accuracy
Resolution Protection
1000
0,1 % R + 8 D
100 m
1414 Vpk
Detection threshold in continuity mode
20
(response time <10ms)
“Active” protection by PTC thermistor
Maximum open-circuit voltage: 3.5V 2V typ
Temperature
8.10.
8.10.1.
Pt100/Pt1000
The user can measure the temperature by means of a Pt100/Pt1000 sensor.
Range
Measurement current
Resolution
Accuracy
Protection
- 125°C à
+ 75°C
< 1 mA
(Pt100)
< 0,1 mA
(Pt1000)
0,1°C
---
0,5 °C
1414 Vpk
- 200°C à
+ 800°C
< 1 mA
(Pt100)
< 0,1 mA
(Pt1000)
0,1°C
---
0,1 % R
1 °C
0,07 % R
1 °C
“Active” protection by PTC thermistor
Display in °C/°F possible
8.10.2.
Thermocouples
Function
Internal temperature
External temperature
Type of sensor
Integrated circuit
K thermocouple
Display range
1000 °C
1000 °F
1000 °C
1000 °F
10 000 °C
10 000 °F
Specified measurement
domain
- 10,0 °C à + 60,0 °C
+ 14,0 °F à + 140,0 °F
- 40,0 °C à + 999,9°C
- 40,0°F à + 1831,8°F
+ 1000 °C à + 1200 °C
+ 1832 °F à + 2192 °F
Uncertainty (note 1)
± 3 °C
± 5,4 °F
1 % R ± 3 °C
1 % R ± 5,4 °F
1 % R ± 3 °C
1 % R ± 5,4 °F
Resolution
0,1 °C
0,1 °F
0,1 °C
0,1 °F
1 °C
1 °F
Thermal time constant (note 2)
0,7 min./ °C
Depending on model of sensor
Detection of sensor open-
circuit
No
Yes: indication of the internal temperature although the external
sensor is connected
Note 1: The stated accuracy in external temperature measurement does not take into account the accuracy of the K
thermocouple.
Note 2: Operation of the thermal time constant (0.7 min/°C):
If there is a sudden variation of the temperature of the multimeter, by 10°C for example, the multimeter will reach 99% of
the final temperature at the end of 5 time constants, or 0.7min/°Cx10°Cx5 cts = 35 min (to which must be added the constant
of the external sensor)
Protection: 1,414 Vpk