- 21 -

Performing a Compression Test

The Chatillon DFX II Series may be used to perform 
a compression test (also called a push test).  A com-
pression test causes a pushing effect on the load 
sensor shaft.  The shaft is commonly equipped with 
a flat adapter or platen.

Normal Mode Test

Place the gauge into Normal mode by pressing the 
PEAK key until the mode indicator about the units 
of measure reads NORM.  Normal mode means the 
gauge will display the real-time load applied to the 
load sensor shaft.  The peak or maximum load is not 
displayed in Normal mode.

Affix your compression test adapter (flat, platen 
fixture, or the like).

Press the Zero key to zero or tare out the weight of 
the adapter or fixture attached to the load sensor 
shaft.

Press the UNITS key to setup your DFX II to display 
load in your required units of measurement.

Apply a pushing force to the load sensor shaft by 
pushing on the sample under tester.  

In a handheld test, make sure to keep the gauge  
perpendicular to the centerline of the test sample 
to avoid side loading effects, e.g. where load is not 
being applied linearly through the load sensor shaft.

In a tester application, where the gauge is mounted 
to a test stand, ensure that the center line of the 
load sensor shaft is aligned directly with the sample 
on the tester.

Observe the load bargraph on the DFX II gauge 
to guard against overloads.  If the bargraph ap-
proached being completely filled, stop the test 
immediately to avoid an overload.

While you are pushing the sample, the compressive 
load indicator will display (converging arrows).

Once the test is completed, press the PEAK key to 
observe the maximum compressive load that was 
achieved during your test.

Peak Compressive Test (C-PK)

This test is performed identically to the Normal 
Mode Test except the gauge is placed in PEAK 
mode.  Placing the gauge in PEAK mode automati-
cally “freezes and displays” the maximum load 
achieved during your test.

1

2

4

Affix Test Adapter

Affix your testing adapter to 
the DFX II load sensor shaft.  
“Finger-tight” the adapter to 
the #10-32 threaded shaft.

Select Normal Mode

Place the DFX II in Normal 
mode by pressing the PEAK 
key until “NORM” is displayed 
above the units indicator.

Select Units of 
Measure

Select the UNITS key to specify 
the units of measure you want 
displayed.

Zero the DFX II

Select the ZERO key to zero 
the DFX II and to tare out the 
weight of the test fixture or 
adapter.

Apply Push Force

Apply a pushing force to the 
sample under test.  Observe 
the load bargraph (fills from 
the center towards the “C”) 
and the Compressive Load 
Indicator (arrows).  The main 
display will indicate the load 
being applied to the load 
sensor shaft using the fixture/
adapter.

Observe Load

Read the force result from the 
display.  If you select the PEAK 
key after you complete your 
test, the DFX will display the 
maximum (Peak) load that was 
achieved during the test.

3

5

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание DFX II Series

Страница 1: ...1 Part No NC003198 July 2013 DFX II Series Digital Force Gauge User Manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...roduct enclosure that might be of sufficient magnitude to cause serious shocks or death Never open the enclosures unless you are an authorized and qualified Chatillon service personnel Never open any...

Страница 3: ...safety requirements of the Low Voltage and the EMC Directives listed and found to be in compliance BS EN 61010 1 2010 Safety Requirement for Electrical Equipment BS EN 61326 1 2006 Electrical Equipme...

Страница 4: ...justment The load bargraph at the bottom of the display indicates the load currently being applied to the internal load sensor The bargraph fills from the center location with tensile loads measured t...

Страница 5: ...ompression Peak C PK Select the Peak key to display the peak load measured in tension or compression Use the zero key to clear the peak reading and to prepare the gauge for the succeeding measure ment...

Страница 6: ...N 100 x 0 1 kgf View Firmware The DFX II Series displays its firmware informa tion To view the firmware characteristics perform the following key sequence 1 INFO Capacity 2 INFO Firmware T C Firmware...

Страница 7: ...rger into the DFX force gauge then plug the battery charger into the correct AC power outlet Ensure the correct power is applied to the battery charger Never plug the USA 2 prong 115VAC charger into a...

Страница 8: ...gauge simply reverse the housing 1 Turn power to the gauge OFF 2 Remove two 2 6 32 screws on the gauge housing using Hex key supplied with the DFX II 3 Rotate the top housing 180 de grees being carefu...

Страница 9: ...ed OFF Auto Shutdown Status PEAK to Change INFO to Proceed ON Auto Shutdown Feature The DFX II Series has an option to automatically shut off power when the gauge is not being used during a 30 minute...

Страница 10: ...o Shutdown 6 INFO Units Lock T C Units Lock Units Lock Status T C Units Lock PEAK to Change INFO to Proceed ON Peak Peak Units Lock Status PEAK to Change INFO to Proceed OFF T LBF 1 23 Norm C Units Lo...

Страница 11: ...ity 2 INFO Firmware 3 INFO Overload 4 INFO Battery 5 INFO Auto Shutdown 6 INFO Units Lock 7 INFO Language Language Peak Lenguaje Peak PEAK to Change INFO to Proceed ENGLISH PEAK para cambio INFO para...

Страница 12: ...et to OFF the Peak Capture is 1000Hz To change the Filter Sample Rate for your DFX II follow this setup procedure and depress the follow ing keys 1 INFO Capacity 2 INFO Firmware 3 INFO Overload 4 INFO...

Страница 13: ...aded load sensor shaft The DFX2 200 has a 5 16 18 male threaded load sensor shaft Be careful to not over tighten adapters as this can damage the load sensor Adapters should be torqued to 5 in lbs 45 N...

Страница 14: ...10 32 SPK FMG 009B 5 16 18 SPK FMG 010A 10 32 SPK FMG 010B 5 16 18 NC000721 5 16 18 NC000725 5 16 18 SPK FMG 012C 5 16 18 SPK FMG 012A 10 32 ML3867 10 32 ML3850 10 32 ML3869 5 16 18 ML3868 5 16 18 17...

Страница 15: ...SPK MT 0001 MT500 SPK MT 0004 LTCM 100 Not Required LTCM 500 Not Required TT NC002582 SPK MT 0001 SPK MT 0004 NC002582 Note These adapters are especially designed to ensure gauge tester centerline ali...

Страница 16: ...your DFX II Series force gauge Customers electing to calibrate their gauge do so at their risk and with the understanding that they are effecting the performance of the gauge The Calibration Procedure...

Страница 17: ...el Peak UNITS to change T C INFO to proceed PEAK to exit Select CALIB Units i UNITS to change INFO to proceed T C DFX II 100 PEAK to go back LBF 0 0 Note The units of measure selected must correspond...

Страница 18: ...force gauge Note Carefully add weights equal to the DFX II gauge s full scale capacity to the hanger fixture Note Verify the compression calibration by checking both the full scale span and zero When...

Страница 19: ...eights but leave tension fixture at tached to the force gauge C C Note Verify the tension calibration by checking both the full scale span and zero When no load is applied the display should read zero...

Страница 20: ...If the bargraph ap proached being completely filled stop the test immediately to avoid an overload While you are pulling the sample the tensile load indicator will display opposing arrows Once the te...

Страница 21: ...rgraph ap proached being completely filled stop the test immediately to avoid an overload While you are pushing the sample the compressive load indicator will display converging arrows Once the test i...

Страница 22: ...SPK FMG 009B Chisel Adapter 100 lbf SPK FMG 008A Chisel Adapter 200 lbf SPK FMG 008B Notch Adapter 100 lbf SPK FMG 010A Notch Adapter 200 lbf SPK FMG 010B Extension Rod 6 inch 10 32 SPK FMG 013A Exte...

Страница 23: ...23 Parts List www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 24: ...24 Parts List www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 25: ...ation hereunder does not include the cost of transportation which must be borne by the customer AMETEK assumes no responsibility for damage in transit and any claims for such damage should be presente...

Страница 26: ...acters Automatic Shut Down 30 minutes May be disabled Power NiMH battery Battery Life Approximately 70 to 100 hours continuous use Instrument Weight 1 5 lbs 0 7 kg Shipping Weights with accessories 4...

Отзывы: