- 18 -
Remove all weight, but
leave tension
fixture attached
T
C
PEAK to go back
LBF
0.0
INFO to proceed
Add full scale
tension load
100 LBF
T
PEAK to go back
LBF
100.0
INFO to proceed
DFX II Series Calibration
Remove all weight, but
leave compression
fixture attached
T
C
PEAK to go back
LBF
0.0
INFO to proceed
Add full scale
compression load
100 lbf
T
PEAK to go back
LBF
00.0
INFO to proceed
Verify compression
calibration
T
C
PEAK to go back
LBF
0.0
INFO to proceed
Place gauge in
tension mode and
exercise 3 times
T
C
PEAK to go back
LBF
100.0
INFO to proceed
Note: Remove all weights but leave compression
calibration hanger attached to the force gauge.
Note: Carefully add weights equal to the DFX II gauge’s
full scale capacity to the hanger fixture.
Note: Verify the compression calibration by checking
both the full scale (span) and zero. When no load is
applied, the display should read zero. When full scale
weights are applied to the fixture, the display should
read the full scale weight equivalent.
Note: Exercise the loadcell three times by pulling on the
loadcell shaft.
C
Note: Carefully add weights equal to the DFX II gauge’s
full scale capacity to the hanger fixture.
Note
Selecting PEAK during the calibration procedure will
take you back to the last step in the procedure.
C
q
q
q
q
Remove full scale
load for zero
verification
T
PEAK to go back
LBF
100.0
INFO to proceed
C
i
i
i
i
i
i
i