background image

- 7 -

View Battery Life

The DFX II Series displays its battery life.  The DFX 
II uses a rechargeable NiMH battery cell. To view 
the DFX II battery life, perform the following key 
sequence:

 

1. INFO <Capacity>

 

2. INFO <Firmware>

 

3. INFO <Overload>

 

4. INFO <Battery>

T

C

Battery

 

Est Battery Life

100 hrs

9.00 Volts

Using The Battery Charger

The DFX II Series may be battery operated or operated  
using the battery charger accessory (SPK-DF2-UNIV).  
When the battery charger is used, the internal  batteries 
are charged.  The battery charger may be plugged into a 
115V or 230V source power outlet.  The battery charger 
contains three interchangeable plug styles: USA, EURO 
and UK.

1. Place the correct plug onto the 
    battery charger.

2. Plug the battery charger into the 
    source power outlet.

3.  Plug the battery charger into 
     the DFX II power input.

4.  Press the DFX II On/Off key to 
     turn power ON.

1

4

3

2

IMPORTANT

Use the Chatillon Battery Charger that comes with 

the DFX II Series digital force gauge.

Do NOT use generic battery chargers.  These can 

permanently damage the gauge electronics.

Note

The DFX II will display “Recharge Battery” when the 

battery voltage drops to approximately 4.5VDC

i

To charge the DFX force gauge plug the phone jack 
of the supplied universal battery charger into the DFX 
force gauge then plug the battery charger into the 
correct AC power outlet.  Ensure the correct power is 
applied to the battery charger.  Never plug the USA 
2-prong 115VAC charger into a 230VAC power source.  
To obtain a full charge it is recommended the DFX force 
gauge is charged for 15-20 hours.  The DFX force gauge 
should get approximately 70 to 100 hours of battery life 
on a single charge.

A battery icon appears on the main operating display 
and operates like a bar graph.  When the gauge has a 
full charge, the battery icon appears full.  As the battery 
voltage diminishes, the battery icon will start to empty.  
You may view the estimated battery life based on cur-
rent usage in hours by depressing the “i” key until the 
battery life screen appears.

The DFX force gauge will alert you when the battery 
assembly needs to be recharged by adding a message 
to the main test screen “Recharge Battery”.  The force 
gauge will shutdown automatically if the battery voltage 
reaches approximately 4.5VDC.

CAUTION: Do not attempt to charge the 
Chatillon DFX series force gauge with any 
other charger than the one supplied with 

the force gauge or supplied by an authorized 

Chatillon dealer.  The nickel metal hydride batteries are 
temperature sensitive and the battery charger is specifi-
cally design to properly charge this battery assembly 
without damaging it.

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание DFX II Series

Страница 1: ...1 Part No NC003198 July 2013 DFX II Series Digital Force Gauge User Manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...roduct enclosure that might be of sufficient magnitude to cause serious shocks or death Never open the enclosures unless you are an authorized and qualified Chatillon service personnel Never open any...

Страница 3: ...safety requirements of the Low Voltage and the EMC Directives listed and found to be in compliance BS EN 61010 1 2010 Safety Requirement for Electrical Equipment BS EN 61326 1 2006 Electrical Equipme...

Страница 4: ...justment The load bargraph at the bottom of the display indicates the load currently being applied to the internal load sensor The bargraph fills from the center location with tensile loads measured t...

Страница 5: ...ompression Peak C PK Select the Peak key to display the peak load measured in tension or compression Use the zero key to clear the peak reading and to prepare the gauge for the succeeding measure ment...

Страница 6: ...N 100 x 0 1 kgf View Firmware The DFX II Series displays its firmware informa tion To view the firmware characteristics perform the following key sequence 1 INFO Capacity 2 INFO Firmware T C Firmware...

Страница 7: ...rger into the DFX force gauge then plug the battery charger into the correct AC power outlet Ensure the correct power is applied to the battery charger Never plug the USA 2 prong 115VAC charger into a...

Страница 8: ...gauge simply reverse the housing 1 Turn power to the gauge OFF 2 Remove two 2 6 32 screws on the gauge housing using Hex key supplied with the DFX II 3 Rotate the top housing 180 de grees being carefu...

Страница 9: ...ed OFF Auto Shutdown Status PEAK to Change INFO to Proceed ON Auto Shutdown Feature The DFX II Series has an option to automatically shut off power when the gauge is not being used during a 30 minute...

Страница 10: ...o Shutdown 6 INFO Units Lock T C Units Lock Units Lock Status T C Units Lock PEAK to Change INFO to Proceed ON Peak Peak Units Lock Status PEAK to Change INFO to Proceed OFF T LBF 1 23 Norm C Units Lo...

Страница 11: ...ity 2 INFO Firmware 3 INFO Overload 4 INFO Battery 5 INFO Auto Shutdown 6 INFO Units Lock 7 INFO Language Language Peak Lenguaje Peak PEAK to Change INFO to Proceed ENGLISH PEAK para cambio INFO para...

Страница 12: ...et to OFF the Peak Capture is 1000Hz To change the Filter Sample Rate for your DFX II follow this setup procedure and depress the follow ing keys 1 INFO Capacity 2 INFO Firmware 3 INFO Overload 4 INFO...

Страница 13: ...aded load sensor shaft The DFX2 200 has a 5 16 18 male threaded load sensor shaft Be careful to not over tighten adapters as this can damage the load sensor Adapters should be torqued to 5 in lbs 45 N...

Страница 14: ...10 32 SPK FMG 009B 5 16 18 SPK FMG 010A 10 32 SPK FMG 010B 5 16 18 NC000721 5 16 18 NC000725 5 16 18 SPK FMG 012C 5 16 18 SPK FMG 012A 10 32 ML3867 10 32 ML3850 10 32 ML3869 5 16 18 ML3868 5 16 18 17...

Страница 15: ...SPK MT 0001 MT500 SPK MT 0004 LTCM 100 Not Required LTCM 500 Not Required TT NC002582 SPK MT 0001 SPK MT 0004 NC002582 Note These adapters are especially designed to ensure gauge tester centerline ali...

Страница 16: ...your DFX II Series force gauge Customers electing to calibrate their gauge do so at their risk and with the understanding that they are effecting the performance of the gauge The Calibration Procedure...

Страница 17: ...el Peak UNITS to change T C INFO to proceed PEAK to exit Select CALIB Units i UNITS to change INFO to proceed T C DFX II 100 PEAK to go back LBF 0 0 Note The units of measure selected must correspond...

Страница 18: ...force gauge Note Carefully add weights equal to the DFX II gauge s full scale capacity to the hanger fixture Note Verify the compression calibration by checking both the full scale span and zero When...

Страница 19: ...eights but leave tension fixture at tached to the force gauge C C Note Verify the tension calibration by checking both the full scale span and zero When no load is applied the display should read zero...

Страница 20: ...If the bargraph ap proached being completely filled stop the test immediately to avoid an overload While you are pulling the sample the tensile load indicator will display opposing arrows Once the te...

Страница 21: ...rgraph ap proached being completely filled stop the test immediately to avoid an overload While you are pushing the sample the compressive load indicator will display converging arrows Once the test i...

Страница 22: ...SPK FMG 009B Chisel Adapter 100 lbf SPK FMG 008A Chisel Adapter 200 lbf SPK FMG 008B Notch Adapter 100 lbf SPK FMG 010A Notch Adapter 200 lbf SPK FMG 010B Extension Rod 6 inch 10 32 SPK FMG 013A Exte...

Страница 23: ...23 Parts List www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 24: ...24 Parts List www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 25: ...ation hereunder does not include the cost of transportation which must be borne by the customer AMETEK assumes no responsibility for damage in transit and any claims for such damage should be presente...

Страница 26: ...acters Automatic Shut Down 30 minutes May be disabled Power NiMH battery Battery Life Approximately 70 to 100 hours continuous use Instrument Weight 1 5 lbs 0 7 kg Shipping Weights with accessories 4...