Introduction 

The Chatillon

®

 DFX II Series digital force gauges is a battery-operated force indicator.  It uses a NiMH bat-

tery for power.  It may also be used with the DFX II battery charger/charger (included) supplied.  

The DFX II Series may be used to measure and display tensile or compressive loads at an accuracy of +0.3% 
of full scale.  The DFX II Series may also be set to capture the peak tensile or compressive load. The DFX Series 
is available in four models having four different capacities and user-selectable units of measurement.

 

Model ozf 

gf 

lbf 

kgf 

N

 

DFX2-010 

160 x 0.2 

5000 x 5 

10 x 0.01 

5 x 0.005 

50 x 0.05

 

DFX2-050 

800 x 1 

25000 x 20 

50 x 0.05 

25 x 0.02 

250 x 0.2

 

DFX2-100 

1600 x 2 

50000 x 50 

100 x 0.1 

50 x 0.05 

500 x 0.5

 

DFX2-200 

200 x 0.2 

100 x 0.1 

1000 x 1

The gauge may be setup to display text in English, French, German, Spanish, Portuguese and Chinese.

Conformance

The Chatillon DFX II Series has been assessed against the essential health and safety requirements of the 
Low Voltage and the EMC Directives listed and found to be in compliance.

 

BS EN 61010-1:2010  Safety Requirement for Electrical Equipment

 

BS EN 61326-1:2006  Electrical Equipment for measurement, control and laboratory use:

   General 

Requirements

 

BS EN 61326-2-3:2006  Electrical Equipment for measurement, control and laboratory use:

 

 

 

Transducers with integrated or remote signal conditioning

The DFX II Series is a RoHS and WEEE compliant device.

Packaging

The Chatillon DFX II Series is supplied with the 
following standard accessories:

l

 

Flat Adapter, 100 lbf (p/n SPK-FMG-011A) or

l

 

Flat Adapter, 200 lbf (p/n SPK-FMG-011B)

l

 

Hook Adapter, 50 lbf (p/n SPK-FMG-012A) or

l

 

Hook Adapter, 100 lbf (p/n SPK-FMG-012B) or

l

 

Hook Adapter, 200 lbf (p/n SPK-FMG-012C)

l

 

DFX II Universal Battery Charger/Charger

   (p/n charger SPK-DF2-UNIV)

l

 

DFX II Carrying Case (SPK-DF-118)

l

 

Certificate of Conformance

Optional Certificate of Calibration with Data is available.

The DFX II is supplied with one (1) flat adapter and  (1) 
hook.  The accessory supplied is dependent on the DFX 
II capacity.

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание DFX II Series

Страница 1: ...1 Part No NC003198 July 2013 DFX II Series Digital Force Gauge User Manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...roduct enclosure that might be of sufficient magnitude to cause serious shocks or death Never open the enclosures unless you are an authorized and qualified Chatillon service personnel Never open any...

Страница 3: ...safety requirements of the Low Voltage and the EMC Directives listed and found to be in compliance BS EN 61010 1 2010 Safety Requirement for Electrical Equipment BS EN 61326 1 2006 Electrical Equipme...

Страница 4: ...justment The load bargraph at the bottom of the display indicates the load currently being applied to the internal load sensor The bargraph fills from the center location with tensile loads measured t...

Страница 5: ...ompression Peak C PK Select the Peak key to display the peak load measured in tension or compression Use the zero key to clear the peak reading and to prepare the gauge for the succeeding measure ment...

Страница 6: ...N 100 x 0 1 kgf View Firmware The DFX II Series displays its firmware informa tion To view the firmware characteristics perform the following key sequence 1 INFO Capacity 2 INFO Firmware T C Firmware...

Страница 7: ...rger into the DFX force gauge then plug the battery charger into the correct AC power outlet Ensure the correct power is applied to the battery charger Never plug the USA 2 prong 115VAC charger into a...

Страница 8: ...gauge simply reverse the housing 1 Turn power to the gauge OFF 2 Remove two 2 6 32 screws on the gauge housing using Hex key supplied with the DFX II 3 Rotate the top housing 180 de grees being carefu...

Страница 9: ...ed OFF Auto Shutdown Status PEAK to Change INFO to Proceed ON Auto Shutdown Feature The DFX II Series has an option to automatically shut off power when the gauge is not being used during a 30 minute...

Страница 10: ...o Shutdown 6 INFO Units Lock T C Units Lock Units Lock Status T C Units Lock PEAK to Change INFO to Proceed ON Peak Peak Units Lock Status PEAK to Change INFO to Proceed OFF T LBF 1 23 Norm C Units Lo...

Страница 11: ...ity 2 INFO Firmware 3 INFO Overload 4 INFO Battery 5 INFO Auto Shutdown 6 INFO Units Lock 7 INFO Language Language Peak Lenguaje Peak PEAK to Change INFO to Proceed ENGLISH PEAK para cambio INFO para...

Страница 12: ...et to OFF the Peak Capture is 1000Hz To change the Filter Sample Rate for your DFX II follow this setup procedure and depress the follow ing keys 1 INFO Capacity 2 INFO Firmware 3 INFO Overload 4 INFO...

Страница 13: ...aded load sensor shaft The DFX2 200 has a 5 16 18 male threaded load sensor shaft Be careful to not over tighten adapters as this can damage the load sensor Adapters should be torqued to 5 in lbs 45 N...

Страница 14: ...10 32 SPK FMG 009B 5 16 18 SPK FMG 010A 10 32 SPK FMG 010B 5 16 18 NC000721 5 16 18 NC000725 5 16 18 SPK FMG 012C 5 16 18 SPK FMG 012A 10 32 ML3867 10 32 ML3850 10 32 ML3869 5 16 18 ML3868 5 16 18 17...

Страница 15: ...SPK MT 0001 MT500 SPK MT 0004 LTCM 100 Not Required LTCM 500 Not Required TT NC002582 SPK MT 0001 SPK MT 0004 NC002582 Note These adapters are especially designed to ensure gauge tester centerline ali...

Страница 16: ...your DFX II Series force gauge Customers electing to calibrate their gauge do so at their risk and with the understanding that they are effecting the performance of the gauge The Calibration Procedure...

Страница 17: ...el Peak UNITS to change T C INFO to proceed PEAK to exit Select CALIB Units i UNITS to change INFO to proceed T C DFX II 100 PEAK to go back LBF 0 0 Note The units of measure selected must correspond...

Страница 18: ...force gauge Note Carefully add weights equal to the DFX II gauge s full scale capacity to the hanger fixture Note Verify the compression calibration by checking both the full scale span and zero When...

Страница 19: ...eights but leave tension fixture at tached to the force gauge C C Note Verify the tension calibration by checking both the full scale span and zero When no load is applied the display should read zero...

Страница 20: ...If the bargraph ap proached being completely filled stop the test immediately to avoid an overload While you are pulling the sample the tensile load indicator will display opposing arrows Once the te...

Страница 21: ...rgraph ap proached being completely filled stop the test immediately to avoid an overload While you are pushing the sample the compressive load indicator will display converging arrows Once the test i...

Страница 22: ...SPK FMG 009B Chisel Adapter 100 lbf SPK FMG 008A Chisel Adapter 200 lbf SPK FMG 008B Notch Adapter 100 lbf SPK FMG 010A Notch Adapter 200 lbf SPK FMG 010B Extension Rod 6 inch 10 32 SPK FMG 013A Exte...

Страница 23: ...23 Parts List www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 24: ...24 Parts List www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 25: ...ation hereunder does not include the cost of transportation which must be borne by the customer AMETEK assumes no responsibility for damage in transit and any claims for such damage should be presente...

Страница 26: ...acters Automatic Shut Down 30 minutes May be disabled Power NiMH battery Battery Life Approximately 70 to 100 hours continuous use Instrument Weight 1 5 lbs 0 7 kg Shipping Weights with accessories 4...

Отзывы: