51
Amplification Curves
Problem
Cause
Solution
2
No amplification
d. Instrument not
calibrated
Calibrate the instrument
e. Dye detector assigned
to incorrect channel
Ensure that the channel
used to detect the
fluorophore matches
correctly
3
No real-time
curve and no
visible bands on
gel
a. Master mix preparation
error
Confirm master mix
calculations
b. Protocol requires
optimization
Test different annealing
temperatures and times
c. Sample component
inhibiting PCR
Dilute sample or change
extraction procedure
d. Excess fluorescent dye
inhibiting PCR
Decrease fluorescent dye
concentration
e. Degraded PCR reagent
Rerun protocol with fresh
reagents