Safety Warnings and Precautions
Warnings and precautions must always be adhered to avoid risk of personal injury and/or
damage to the instrument. If the instrument is used in a manner not specified by the
manufacturer, the protection provided by the instrument may be compromised. The information
in this manual is intended to supplement the normal safety requirements established in the
user’s country.
The Open qPCR instrument complies with safety standards IEC 61010-1, level of pollution 2, and
overvoltage category II.
General Safety
The Open qPCR Instrument must be connected to a 3-wire grounded
outlet with the correct voltage rating. A break in the electrical ground
path may create a hazardous condition.
Do not touch the power switches or power cord with wet hands.
Always power down the instrument and disconnect the power cord before
performing cleaning or maintenance procedures.
The heat block and lid temperature rises above 95
°
C. To avoid injury, do
not touch the heat block, lid, or tubes immediately after a run. Let the
surface cool for a few minutes before removing the tubes. If samples
must be removed immediately, avoid contact with surfaces by extracting
tubes with tweezers.
Do not deliberately modify safety features of the instrument.
Do not block ventilation at the front and back of the instrument. Leave at
least six inches of space on all sides of the instrument when powered on.
Wear protective eyewear, clothing, and gloves when handling reagents or
operating the instrument.