Figure 3.18
11.
If calibration fails, you will see a screen displaying
Calibration Failed
(Figure 3.19).
Repeat the calibration process in such instances. If failure persists on repeat, contact
Technical Support.
Figure 3.19
It is not recommended to reuse the calibration dyes because photo-bleaching occurs at
different levels from well to well. If you would still like to reuse the dyes, pool the
solution first and aliquot again as before.