Amplification Curves
Problem
Cause
Solution
6
Jagged curves
a. Very low fluorescence
signal
Optimize fluorescent dye or
probe concentrations
b. Air bubbles present or
droplets adhered to tube
walls
Spin down tubes for 10
seconds using a
microcentrifuge
c. Incorrect tube type
used
Use Chai's validated PCR-
strip tubes. The instrument
requires 100
µ
L optically
clear, non-auto fluorescent
tubes
7
Very low
fluorescence
signal
a. Low copy number of
template
Increase concentration or
add more volume of sample
b. High background signal
Decrease fluorescent dye or
probe concentration. Check
probe purity
c. Sample component
inhibiting PCR
Dilute sample or change
extraction procedure
d. Degraded PCR reagent
Rerun protocol with fresh
reagents