Fig.19 Stopper of main control panel
4.9 Light Path
In case of the significant inclination of fiber image in X/Y camera, the position of CMOS camera plate of
X/Y camera may be slightly changed due to the long-term vibration or self-stress. To rectify the fault,
unscrew the screws on the camera plate of X/Y camera (as shown in Fig.20), manually adjust the
position of camera plate of this image and observe the fiber imaging condition in the horizontal
double-vision video of X/Y camera, and fix the screws when the image is centered and flush with another
image.
In case of dark spots or shadows on the video background of X/Y camera, there may be the dirt on the
X/Y camera microscope lens or X/Y camera CMOS surface To rectify the fault, wipe the surface of front
lens of microscope with a cotton swab soaked in alcohol, and quickly blow it dry with ear washing bulb.
At this time, close the windshield and observe the screen imaging. If the dark spot still exists on the X/Y
camera, loosen the screws of camera plate, take off the camera plate, gently wipe the CMOS surface
with a cotton swab soaked in alcohol, blow it dry with the side of ear washing bulb, re-adjust the camera
plate to make X/Y fiber imaging at center and vertically aligned.
Stopper
fixing
the
control board
X26 cable
X4 socket
Содержание 6481 Series
Страница 8: ...6 Fig 4 Jam of propel motor...
Страница 11: ...9 Fig 8 Align motor Screws of align motor...
Страница 14: ...12 Fig 11 Imaging lamp Screws fixing the imaging lamp Cable at spindle may be damaged Weld point may be broken...
Страница 16: ...14 Fig 13 Replacement of spindle...
Страница 21: ...19 Fig 18 Main control panel FPGA load indicator V17 X4 socket...