Fig.14 Installation of imaging lamp
4.5 Heater
If the heater does not work or can not automatically heat (automatic heating function is on in the menu),
first check whether cables X11 and X12 of control board are plugged in. If they are plugged in, it means
that the heater is damaged and shall be replaced. Unplug cables X11 and X12, and remove the fixing
screws of heater (Fig. 15) to replace the heater.
Fit the cable
through here