Fig.9 Align micrometer head
4.4 Imaging Lamp
White screen, sometimes with the discharge, occurs at the startup, or the instrument restarts when the
windshield is opened or closed, or X/Y camera looks like black screen after the windshield is closed, it
may be that the imaging lamp cable is damaged at the spindle or the imaging lamp cable is broken at the
weld point of printed board, the imaging lamp and spindle shall be replaced.
①
As shown in Fig.10, remove the limiting plate of sketch mounting plate to present the imaging
lamp, and remove 4 screws fixing the imaging lamp (Fig.11).
②
As shown in Fig. 12, pull out X26 plug and jet out X26 pin from plug with tweezer and do not
discard the plug.
③
Pull out the X26 cable from the spindle, and cut it directly if it is not easy to pull out.
④
As shown in Fig. 13, remove the spindle with a flat-blade screwdriver and replace it with a new
one.
Align micrometer head