5B-10
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Dual Connector Cable
IUI Board
Snap Rivet
(do not cut)
Flex
Cable
Syringe Size Sensor
IUI Support Bracket
Retainer Clip
5.2.7
Removing IUI Board Assembly
1. If present, remove Retainer Clip from
Dual Connector Cable-IUI Board
connection.
2. Disconnect Dual Connector Cable from
IUI
Board
J2
.
3. Disconnect Syringe Size Sensor
harness from
IUI
Board
J4
.
4. Disconnect Flex Cable from
IUI
Board
J3
.
5. Use small diagonal cutters to lift and
remove snap rivet from
IUI
Board.
6. Remove
IUI
Board.
During Reassembly:
Route Dual Connector Cable and Syringe
Size Sensor harness under
IUI
Support
Bracket.
Do not cut rivet.
CAUTION