5B-25
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.21 Removing Leadscrew Assembly and
Bottom Plate
1. Remove two (
2
) screws attaching
Leadscrew Assembly to Bottom Plate.
2. Remove Leadscrew Assembly.
3. Remove remaining screw attaching
Bottom Plate to Extrusion Frame.
4. Remove Bottom Plate.
During Reassembly:
If replacing Bottom Plate, remove bushings
and grommets from old plate and install into
new plate.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Leadscrew Assembly
Bushing (8 PL)
Bottom Plate
Extrusion Frame
Grommet (4 PL)