5B-6
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
5.2.4
Removing Door Assembly
1. Remove four (
4
) T8 Torx Security screws
and four (4) Hinge Bushings attaching
Door Assembly to right side of Front
Case.
2. Unlock Door, and lift Lock Assembly
away from Door. Take care not to pull
on Lock Assembly harness.
3. Remove Door Assembly.
Disassembling Door Assembly:
4. Remove four (
4
) screws attaching Front
Door, Rear Door Lockplate, and Rear
Door.
5. Remove seven (
7
) screws attaching
Hinge to Front Door and Lockplate.
6. Remove two (
2
) screws attaching
Lockplate and Strikeplate Lockplate to
Front Door.
During Reassembly:
Reassembling Door Assembly:
1. Use two (
2
) screws to assemble Door
Strikeplate to Front Door.
2. From inside of Door, align Front Door
Lockplate holes with through-holes on
Door. From outside of Door, align holes
on hinge with those on Door and Front
Door Lockplate. Install but only loosely
tighten seven (
7
) screws to attach hinge.
Front Case
Door
Assembly
Lock
Assembly
Harness
T8 Torx Security
Screws and Hinge
Bushings (x4)
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......