Table 6-4. Error Codes
Refer to the PC Unit / Pump Module (
8000
/
8100
Series) Technical Service Manual for details on
accessing the Maintenance Mode and viewing the Error Log.
TROUBLESHOOTING
6-7
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
350.6600
Logic Board
Watchdog circuit failed power-
on self test.
If error is repeatable: Replace
351.6660
Linear Position Sensor
Movement error; predicted
change in linear position
sensor inconsistent with
expected position.
If error is repeatable: Replace
linear position sensor.
350.6630
Force Sensor
Pressure not tracked by force
sensor.
If error is repeatable: Replace
351.6610
Motor Watchdog
Unexpected countdown
interrupt.
If error is repeatable: Replace
351.6720
Motor
Logic Board
Motor steps accumulating too
fast or too slow.
If error is repeatable: Replace
351.6730
Motor
Logic Board
Too many steps in motor step
sample.
If error is repeatable: Replace
351.6740
Engagement Nut
Nut not engaged during
infusion; lever not opened.
If error is repeatable: Replace
Logic Board, and/or nut
engagement sensor or
351.6750
Engagement Nut
Attempt to engage nut timed
out before engaging.
If error is repeatable: Replace
Logic Board, and/or nut
engagement sensor or
351.6760
Module
Calibration flag not set.
Module not calibrated.
Calibrate, as required.
350.6680
Logic Board
Timer bases; clock frequency
out of sync.
If error is repeatable: Replace
Error Code
Subsystem
Explanation
Response
352.6640
Force Sensor
Force sensor output is zero or
is outside acceptable range.
If error is repeatable: Replace
351.6770
Pressure Sensing Disc
Logic Board (Syringe Module
only)
Pressure sensing disc circuit
test failed during power-on
self test.
If error is repeatable: Replace
pressure sensing disc and/or
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......