1. After accessing
System Config -
Syringe
options display, press
All Mode
soft key.
2. To change option setting, press soft key
next to applicable option (
Enable
or
Disable
).
• Option setting changes and
System
Config - Syringe
page
1
displays.
2.4.2
Auto Pressure
Enabling
Auto Pressure
allows the Auto
Pressure option to be displayed in the
Pressure Limit screen when a pressure
sensing disc is in use. Auto Pressure
automatically sets the alarm limit for a
shorter time to alarm, as follows:
• If current pressure is
100
mmHg or less,
system adds
30
mmHg to current
pressure, to create a new alarm limit.
• If current pressure is greater than
100
mmHg, system adds
30%
to current
pressure, to create a new alarm limit.
1. After accessing
System Config -
Syringe
options display, press
Auto
Pressure
soft key.
2. To change option setting, press soft key
next to applicable option (
Enable
or
Disable
).
• Option setting changes and
System
Config - Syringe
page
1
displays.
2.4.3
Back Off
When
Back Off mode
is enabled and a
pressure sensing disc is in use, the motor
reverses plunger movement during an
occlusion until the pressure returns to
preocclusion levels, automatically reducing
bolus flow.
1. After accessing
System Config -
Syringe
options display, press
Back Off
soft key.
2. To change option setting, press soft key
next to applicable option (
Enable
or
Disable
).
• Option setting changes and
System
Config - Syringe
page
1
displays.
CHECKOUT AND CONFIGURATION
2-4
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
2.4
CONFIGURATION SETUP - SYRINGE
MODULE
(Continued)
2.4.1
All Mode
(Continued)
System Configuration - Syringe
Back Off mode
Enable
Disable
Allows use of the
back off feature.
System Configuration - Syringe
Auto Pressure
Enable
Disable
Allows use of the
auto pressure feature.
System Config - Pump 2 of 3
System Configuration - Syringe
All mode
Allows use of the
All mode feature.
Enable
Disable
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......