Chapter 2 — CHECKOUT AND CONFIGURATION
2-1
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
2.1
INTRODUCTION
This chapter describes initial setup and
configuration for the Syringe and PCA
Modules.
NOTE:
Due to product changes over time,
configurations described in this chapter may
differ from the instrument under service. If
there are any questions, contact CareFusion
Technical Support.
2.2
NEW INSTRUMENT CHECKOUT
Prior to placing a new instrument in use,
perform a check-in procedure using the
Maintenance Software.
When powering up the instrument, verify the
instrument beeps and all display
LED
segments flash. This confirms that the
instrument has performed its self test and is
operating correctly. During operation, the
instrument continually performs a self test,
and will alarm and display a message if it
detects an internal malfunction.
Contact CareFusion authorized service
personnel if the instrument has physical
damage, fails to satisfactorily pass the
startup sequence, fails a self test, or
continues to alarm.
Should an instrument be jarred severely or
dropped, remove it from use immediately. It
should be thoroughly tested and inspected by
qualified service personnel to ensure proper
function prior to reuse.
CAUTION
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......