CORRECTIVE MAINTENANCE
Syringe Module
5A-14
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.12 Removing Top Disk Holder
1. Remove two (
2
) screws attaching Top
Disk Holder to Front Case.
2. Remove Top Disk Holder.
NOTE:
When replacing the Top Disk Holder,
remove the Flag Disk Holder and Spring from
the old part and reuse.
During Reassembly:
• Set spring torque by inserting end of
Spring in hole of Flag Disk Holder and
winding holder counterclockwise.
• Ensure Silicone Tubing and Gasket in
Top Disk Holder mounting area on Front
Case are in place and undamaged. (See
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Flag Disk Holder
Spring
Top Disk Holder
To replace “Older Version” Top Disk Holder
(see below), replace with Front Case
9
).
CAUTION
“Older Version”
Top Disk Holder
See
CAUTION
statement above.
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......