5.2.3
Removing Handle
1. Remove six (
6
) screws attaching Left
and Right Handles.
2. Remove two (
2
) screws attaching Left
Handle to Lock Housing.
3. Detach Left and Right Handles from
each other and module.
During Reassembly:
• Assemble Left Handle over Lock Housing
and ensure that Lock Harness is routed
under Lock Housing boss.
• Use two (
2
) screws to attach Left Handle
to Lock Housing.
• Install Right Handle over Left Handle and
secure with six (
6
) screws.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
5B-5
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
Right Handle
Left Handle
Rear Case
Screws attach Left Handle
to Lock Housing
Lock Harness
Lock Housing