TROUBLESHOOTING
6-8
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
354.6690
Logic Board
Pressure Sensor current
monitor: current is outside
acceptable range.
If error is repeatable: Replace
Logic Board or Pressure
Table 6-4.
Error Codes
(Continued)
Error Code
Subsystem
Explanation
Response
352.6700
Logic Board
Force Sensor
Force sensor circuit check
failure.
If error is repeatable: Replace
Logic Board or Force Sensor
356.6710
Digital Sensor
Digital sensor is outside
acceptable range during no-
power test.
If error is repeatable: Replace
digital sensor.
353.6650
Linear Position Sensor
Linear position sensor output
is zero or is outside
acceptable range.
If error is repeatable: Replace
linear position sensor.
355.6620
Syringe Sizer Sensor
Syringe sizer output is outside
acceptable range.
If error is repeatable: Replace
syringe sizer sensor.
354.6670
Pressure Sensing Disc
(Syringe Module only)
Pressure sensor output is
outside acceptable range.
Replace pressure sensing
disc.
Replacing the Force Sensor Assembly or Logic Board Assembly in a module with version
7
or earlier software
356.6780
PCA Handset
System detected PCA
Handset Key stuck at power-
up.
If error is repeatable: Replace
PCA Handset.
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......