CANON Digital Galvano Scanner System
KP-1SM GC-251
Users Manual 1.00
79
Parameter ID
51
Parameter
Name
Encoder VR Adjustment
Head 2 A/B-phase Offset
Data
A/B-phase offset of encoder VR adjustment head 2 (Unit: none)
Return Value
0: Parameter setting successful
1: Parameter setting unsuccessful
Explanation
This parameter is used to adjust the offset of encoder signal.
Note: This is usually set appropriately at shipping.
Related
Command
/Parameter
Parameter ID
52
Parameter
Name
Encoder VR Adjustment
Head 2 A-phase Amplitude
Data
A-phase amplitude of encoder VR adjustment head 2 (Unit: none)
Return Value
0: Parameter setting successful
1: Parameter setting unsuccessful
Explanation
This parameter is used to adjust the A-phase amplitude of the encoder signal.
Note: This is usually set appropriately at shipping.
Related
Command
/Parameter
Parameter ID
53
Parameter
Name
Encoder VR Adjustment
Head 2 B-phase Amplitude
Data
B-phase amplitude of encoder VR adjustment head 2 (Unit: none)
Return Value
0: Parameter setting successful
1: Parameter setting unsuccessful
Explanation
This parameter is used to adjust the B-phase amplitude of the encoder signal.
Note: This is usually set appropriately at shipping.
Related
Command
/Parameter