Table 7-7: CSAT3H heater controller outputs
Variable name
Units
Description
mV_real_arms_Max
mV
Maximum millivolt signal CSAT3H heater controller used
to determine how much heat to apply to the arms
wnd_spd_fctr_Max
–
Maximum multiplier applied to heaters as conditions
become more windy, requiring more heat (value of 0 to
7)
T_boost_mult_Max
–
Maximum multiplier applied to heaters under certain
conditions that require more heat (value of 0 to 16)
T_Panel_CR300_Avg
°C
Average internal temperature of the CSAT3H heater
controller
volt_batt_CR300_Min
V
Minimum battery voltage supplied to the CSAT3H heater
controller
8. Maintenance and
troubleshooting
8.1 General maintenance
With no moving parts, maintenance of the CSAT3BH is minimal and limited to the following:
l
Replacing the desiccant canister
l
Monitoring diagnostics and measurement offsets to determine when factory recalibration
is needed
l
Monitoring diagnostics to assure battery bank for both 12 and 24 VDC power sources have
sufficient charge
l
Replacing the desiccant pack within the CSAT3H heater electronics enclosure annually
The following sections address these maintenance activities.
CSAT3BH Three-Dimensional Heated Sonic Anemometer