FIGURE 6-22. RS-485 cable connections
6.10.3 USB
If data collection from the anemometer is to be done by a computer using USB communications,
first connect to the CSAT3BH as described in
(p. 24) to confirm the following three
settings:
Operating Mode (choose one of the following):
l
Mode 2: Self-triggered | No filter | Unprompted output to computer
l
Mode 3: Self-triggered | Filtered | Unprompted output to computer (if Mode 3, select
filter bandwidth)
Unprompted Output Port: USB Port
Unprompted Output Rate: 10, 20, 50, or 100 Hz
NOTE:
In
Device Configuration Utility
, the settings for Unprompted Output Port and Unprompted
Output Rate will be grayed-out until the operating mode has been set to Mode 2 or Mode 3.
To complete installation, connect a CSAT3BCBL2 to the Power/SDM port by screwing the M16
connector into the port until tight.
CSAT3BH Three-Dimensional Heated Sonic Anemometer
59