Table 7-7: CSAT3H heater controller outputs
Variable name
Units
Description
delta_T_trnsds_Max
°C
Difference between the maximum temperature of top
and bottom transducers
T_arm_uppr_Avg
°C
Average temperature of the top sonic arm
T_arm_uppr_Std
°C
Temperature standard deviation of the top sonic arm
T_arm_lwr_Avg
°C
Average temperature of the bottom sonic arm
T_arm_lwr_Std
°C
Temperature standard deviation of the bottom sonic arm
T_arm_Max
°C
Maximum temperature between the top and bottom
arms
T_arm_Min
°C
Minimum temperature between the top and bottom
arms
delta_T_arms_Max
°C
Difference between the maximum temperature of top
and bottom arms
T_trnsd_arm_Max
°C
Maximum temperature among all transducer and arm
thermistors
set_T_pt_trnsds_Avg
°C
Temperature to which transducers are set to heat; NAN if
no heating is needed
set_T_pt_arms_Avg
°C
Temperature to which arms are set to heat; NAN if no
heating is needed
T_amb_ctrl
°C
Average ambient temperature measured by the temp/RH
probe connected to the CSAT3H heater controller
T_amb_ctrl_Std
°C
Ambient temperature standard deviation measured by
the temp/RH probe connected to the CSAT3H heater
controller
RH_amb_ctrl
%
Average ambient relative humidity measured by the
temp/RH probe connected to the CSAT3H heater
controller
T_DP_amb_ctrl
°C
Average ambient dew point temperature determined by
the temp/RH probe connected to the CSAT3H heater
controller
CSAT3BH Three-Dimensional Heated Sonic Anemometer
80