5. Humidity values in practice
CAMAG has performed several tests using potassium acetate,
potassium thiocyanate and sodium chloride.
Our experiments
have shown that an average humidity of approx. 47% can
be achieved successfully with potassium thiocyanate
. The
humidity values measured by the ADC2 where generally slightly
below the theoretical values: for 47% rH, 44 - 47% rH and for
75% rH, 69 - 73% rH was measured. The reason why the theo-
retical humidity cannot be fully reached is due to the adiabatic
process and the energy balance during aggregate state
changes. A good example is shown in the graphics below:
28
30
32
34
36
38
40
42
44
46
48
0
1
2
3
4
5
[% RH]
[Minutes]
RH
°C
Fig.16: Measurement of the relative humidity in the ADC2 at start of the humidity con-
trol. Used salt: Potassium thiocyanate, relative humidity in the lab approx. 29%.
22.5
23
23.5
24
24.5
25
0
1
2
3
4
5
[°C]
[Minutes]
RH
°C
Fig.17: Measurement of the temperature in the ADC2 at start of the humidity control.
Used salt: Potassium thiocyanate, temperature in the lab: approx. 23°C. The higher
temperatures in the instrument at start are caused by warm up / non equilibrium ef-
fects.