5.3. Procedure for saving a changed method:
9
Select the parameter group
Save method
or press the
END
key after having changed the parameters.
9
Select the memory position to save the new method by
means of the
5
or
6
key. Press
ENTER
to confirm position
and save the parameters.
5.4. Local delete of a saved method:
9
Press the
RUN
key.
The most recently used method is displayed. With the
5
or
6
key you can select the locally stored method to delete.
9
Press the key combination
RUN-END
(press and hold
RUN
,
then press
END
) and confirm deleting with
ENTER
.
)
Flow diagrams for these dialogs are depicted in chapter 6.
5.5. Aborting an application in progress
Procedure:
9
Press the
END
key
The development in progress is aborted and if the plate was
immersed in solvent, a drying process will follow. An aborted
development cannot be continued.
5.6. Resetting the instrument
Procedure:
9
Press the
RESET
key
This will abort all activities in progress and re-initialize the
instrument.
Manual, August 2005