ConneX1 Operator’s Manual
30
Performing a Manual Bump Test
Depending on the setpoints, the low or high alarm displays.
6.
When the following screen displays, close the valve on the gas cylinder
and remove the calibration cap from the detector.
7.
When calibration is complete, the following screen displays.
8.
The LCD returns to the Bump Test Configuration Menu screen.
If the bump test failed, repeat the bump test again or refer to
18. Troubleshooting
on page 40
.
11. Datalogs
The detector records datalog samples that can be compiled to create a report
using Fleet Manager II.
Using Fleet Manager II, define the logging interval from 1 to 600 seconds. Default
is 5 seconds.
The total number of 8-hour days of datalogs that can be recorded, assuming 90%
of the day has no gas concentrations.