ConneX1 Operator’s Manual
13
General User Options
Distributor Name:
Name of the distributor company.
Enable Force Bump:
If Force Bump is enabled and the sensor is past due for a
bump test, the following screen displays during the startup self-tests.
(Note: “ To start” alternates with the progress bar every 4 seconds.)
The sensor must be bump tested to continue and enter normal operation. Press
, and refer to
9.2 Performing a Manual Bump Test on page 29
.
Or
Press and hold
to deactivate the detector.
A value must be entered in the
$ &
field in the Sensor
Configuration section before enabling
Force Bump
.
CAUTION
!"
option is automatically disabled.
NOTE
BW Technologies by Honeywell recommends to bump test the
#
to gas by exposing the detector to a gas concentration that
exceeds the alarm setpoints. Verify that the audible and visual
alarms activate. Calibrate if the readings are not within the
For complete instructions to perform a bump test, refer to
9.2 Performing a Manual
Bump Test on page 29
.
The detector is shipped with the
Force Bump
option disabled.
Enable Force Calibration:
If enabled and a sensor is past due for calibration or
the last calibraiton failed, the following screen displays during the startup self-tests.
The sensor must be calibrated to continue and enter normal operation. Press ,
and refer to
10.4 Manual Calibration Procedure on page 27
.
Or
Press and hold to deactivate the detector.
Содержание CONNEX1
Страница 1: ...Gas Detector Operator s Manual...
Страница 2: ......
Страница 7: ...ConneX1 Operator s Manual v...
Страница 8: ...ConneX1 Operator s Manual vi...
Страница 9: ...ConneX1 Operator s Manual vii...
Страница 71: ......
Страница 72: ...D6657 0 50104992 032 BW Technologies by Honeywell 2012 All Rights reserved...