Bump Test Using the IntelliDoX
10. Bump Test
A bump test is the process of applying a small amount of test gas to force the
detector into alarm. A bump test should be performed daily to confirm the sensor is
responding correctly to gas, and that the audible, visual, and vibrator alarms activate
during an alarm condition. Calibrate if the readings are not within specified limits.
10.1 Bump Test Using the IntelliDoX
The IntelliDoX can peform bump tests. Refer to the IntelliDoX Technical Reference
Guide for more information.
10.2 Performing a Manual Bump Test
CAUTION
BW Technologies by Honeywell recommends to bump test the sensor
#&
exposing the sensor to a gas concentration that exceeds the alarm
setpoints.
Follow this procedure when
Force Bump
is enabled and a bump test is required
during startup. To perform a bump test with the IntelliDoX, refer to the IntelliDoX
Technical Reference Guide.
1.
Connect the calibration hose to the 0.3 l/min regulator on the gas cylinder.
2.
Connect the calibration hose to the intake inlet on the calibration cap. Refer
to
11.3 Installing the Calibration Cap on page 26
.
3. Press
to enter Configuration Mode. Scroll down to
Bump Test
. Press .
4.
When the following screen displays, attach the calibration cap and apply
bump test gas at a flow rate of 250-500 mL/min. Refer to
10.3 Installing
the Calibration Cap on page 26
.
5.
As the sensor detects gas, it will go into alarm.
Содержание CONNEX1
Страница 1: ...Gas Detector Operator s Manual...
Страница 2: ......
Страница 7: ...ConneX1 Operator s Manual v...
Страница 8: ...ConneX1 Operator s Manual vi...
Страница 9: ...ConneX1 Operator s Manual vii...
Страница 71: ......
Страница 72: ...D6657 0 50104992 032 BW Technologies by Honeywell 2012 All Rights reserved...