16
2.
With covers removed, use a 3mm Allen key to remove the rotational stop screw (
B
)
and lock washer (
C
) from the back of spring arm post. Take care to ensure the lock
washer (
C
) and shim washer (
A
) do not fall off during this step. See
Figure 19
.
3.
Connect the spring arm electrical connector to the mounting post electrical connector
by pushing them together.
4.
Slide the spring arm into the mounting post. Ensure the washer (
A
) is present and did
not fall off the spring arm shaft.
5.
Using 3mm Allen key, fasten the rotational stop screw with lock washer into
the back
of the spring arm post. Refer to
Figure 20
.
Figure 19: Remove stop screw
.
Figure 20: Install stop screw
.
WARNING: Always make sure the rotational stop screw is properly fastened
to avoid a safety hazard.
6. Gently snap right hand cover onto spring arm in reverse order of removal instructions
described above. When performing this step, ensure that the sliding cover (already on
the spring arm) mates with a groove on the inside of the spring arm cover. Refer to
Figure 21.
7.
Gently snap left hand cover onto spring arm in reverse order of removal
instructions described above. When performing this step, ensure that the sliding cover
(already on the spring arm) mates with a groove on the inside of the spring arm cover.
Also ensure that both covers mate properly to each other as they are snapped
together. Refer to Figure 22.
Sliding
Cover
Figure 21 & 22: Left
and right hand cover
assembly
.
For instruction on how to Mount the Light Head to the Spring Arm follow steps in section
“
9.0
Mounting the Light Head to the Spring Arm
”
Содержание AIMLED070 Series
Страница 36: ...36 10 0 Equipment Anchorage Diagrams Figure 46 Anchorage Seismic single ceiling model ...
Страница 37: ...37 Figure 47 Anchorage Seismic double ceiling model ...
Страница 38: ...38 Figure 48 Anchorage Seismic wall model ...
Страница 39: ...39 Figure 49 Anchorage Seismic wall model ...
Страница 42: ...42 13 0 Equipment Maintenance Log ...
Страница 43: ...43 NOTES ...