50
MENU 3B
1
st
digit: Inhibit the setpoint "0", enable the setpoint "1" or enables the lacht
setpoint (latch) "2".
2
nd
digit: Activation mode HIGH "0" or LOW "1".
3
rd
digit: Delay the relay activation by timer"0", asymetrical hysteresis(HYS-1)
"1" or asymetrical hysteresis (HYS-2) "2".
4
th
digit: Activation by fast value "0", by net value "1", by gross value "2", by
peak value "3", by valley value "4", by peak to peak value "5" .
5
th
digit: Output normaly open "0", output normaly closed "1".
B
A
TARE
HOLD
LIMIT
MAX
DSP2
INP2
DSP1
INP1
TARE
RESET
LIMIT
ESC
PROG
TEACH
RUN
MIN
DATA
FLT
STORE
2
1
4
3
MAX/MIN
DATA
ENTER
Digit 1
0=OFF
1=ON
2=ON (latch)
Digit 2
0=HI
1=LO
Digit 3
0=Delay
1=Hyst_1
2=Hyst_2
Digit 4
0=Fast
1=Net value
2=Gross value
3=Peak value
4=Valley value
5=Peak to peak value
Digit 5
0=Normaly open
1=Normaly closed
NEW FUNCTIONS
burster ISO
FUNCTION
Answer
type
I
0I
Transmits logical input status
Y
0Y
Transmits Peak to peak value
Returns value
n
0n
LATCH relays reset
y
0y
Peak to peak
Only command, no answer
In Page 21 (RS2) or page 22 (RS4) there is a new protocol 3 = MODBUS (See MODBUS MANUAL Edition JUNE 2000 COD.
30727077).
PROGRAMMABLE FUNCTION nº 31
Fast transmmision via RS232C or RS485
During the activation of logical function nº31 the instrument will send through serial output RS2 or RS4 at 200/s the value
measured, with the choosed tansmission format.
message format:
polarity
X X X . X CR
The decimal point position is an example. It could be in any position.
This function allows data capture in a file for after analisis by software (example EXCEL).
burster Protocol :
1 start bit, 8 bit data, no parity, 1 stop bit.
ISO1745 Protocol: 1 start bit, 7 bit data, 1 bit parity even parity, 1 stop bit.
56