Connect the instrument to the main supply. During a few 
seconds  al  segments  will  turn  on  as  a  checking  of  the  good 
function of the device. 
Press 

ENTER

 key to enter to the programming mode and the 

display will show the indication -Pro-. 
The programming rutine is divided in independant access 
modules, accessible by pressing the 

 key from the level 

–pro- in this order: 
  
1.

 

CnInP =Input configuration. 

2.

 

CndSP =Display configuration. 

3.

 

SetP =Setpoints. 

4.

 

Anout =Analog output. 

5.

 

rSout = Serial output. 

6.

 

LoGIn =Logical input. 

The 3, 4 and 5 modules will be omitted if

 

the 2RE option, 

analog option or serial option are not installed.

 

The 

information about the programming of each one is explained 
in its manuals.

 

The picture shows how to enter in the programming mode, 
the module selection level and the exit saving changes or not 
saving changes. Once the display shows the indication of the 
module, the access to the configuration menus is done by 
pressing

 

ENTER

 

key. 

 

The 

ESC

 key exits the programming mode from any 

programming step. 
 
The global diagrams, that can be seen in the picture, shown 
the way to advance in the programming of the instrument 

 

2.2 – Programming instructions

 

3.3 - Table of programmable functions  

 

 

Nº : Function number. 

 

Function : Function name. 

 

Description : Description and characteristics of the function. 

 

Activation :  

 

Falling edge : The operation is performed on a falling edge applied to the pin with respect to COMMON. 

 

Low level : The function remains activated while the corresponding pin is held at a low level with respect to COMMON. 

-

 

(*) Default factory configuration. It can be restored by programming all pins to '0'. 

 

(1) Activating the functions 3 and 4 simultaneously the peak-peak value is shown. 

Nr Function 

Description 

Activation 

0 Deactivated 

None 

None 

TARE (*) 

Adds the current display value to the tare memory. The display goes to zero 

Pulse 

RESET TARE 

Adds the tare memory contents to the display value and clears the tare memory   Pulse 

PEAK    (1) 

Display PEAK value. 

Fixed level 

VALLEY  (1) 

Display VALLEY value. 

Fixed level 

RESET PEAK/VALLEY 

Clears the peak or valley memory (if the values are on display) 

Pulse 

6 PEAK/VALLEY 

(*) 1

st

 push recalls peak, 2

nd

 push recalls valley, 3

rd 

push recalls peak-peak, 4th 

push brings the meter to the indication of the variable being measured 

Pulse 

RESET (*) 

Combined with (1) delete the tare. 
Combined with (6) delete peak or valley.  

Pulse combined with (1) 
or (6)   

HOLD1 

Holds the display while the outputs remain active 

Fixed level 

HOLD2 (*) 

Holds the display, the RS and the analog outputs 

Fixed level 

0 to 9 : DISPLAY / MEMORY FUNCTIONS 

10 to 12 : FUNCTIONS ASSOCIATED WITH THE DISPLAY OF THE INPUT VARIABLE

Nº Function 

Description 

Activation 

10  INPUT 

Displays the actual input signal value in mV (flashing) 

Fixed level 

11  GROSS 

Displays the measured value + the tare value = gross 

Fixed level 

12  TARE 

Displays the amount of tare contained in the memory 

Fixed level 

31

Содержание 9181

Страница 1: ...span are 9999 and 9999 points a To validate data and advance to the next point press ENTER or b To save the programmed data in the memory and return to the run mode the meter is scaled by three point...

Страница 2: ...E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN 2 INTRODUCTION TO THE DIGIMASTER This manual does not constitute a formal agreement All information given in th...

Страница 3: ...ond the indication P 30 appears in the display 3 23 23 2 Input 30 value B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG T...

Страница 4: ...lter The figure 24 2 shows the initially selected level for the filter P any number between 0 and 9 with the FLT LED activated Press repeatedly the key to change the digit until desired value appears...

Страница 5: ...mA or lineal potentiometer for measure distance length etc The instrument supply different excitation voltages for sensors 24V 30mA 2 2V 30mA 5 or 10V 120mA allowing supply up to 4 Load Cells This dif...

Страница 6: ...values 26 This menu allows selection among 4 levels of display rounding When resolution is not critical a rounding increment higher than 1 may help to stabilize the display MENU 2AB ROUND FILTER Progr...

Страница 7: ...Indicates exit from de program mode with data memory storage 27 This menu allows enabling and disabling the tare function and its reset MENU 2 TARE The initially programmed option appears on the displ...

Страница 8: ...value different from zero is contained in the tare memory To reset the tare memory press and hold down the RESET key then press the TARE key Release first TARE then RESET To take a tare or reset it ba...

Страница 9: ...d VALEY distance value detected since the last reset or tare B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G...

Страница 10: ...m wiring connections remove the terminal block from the meter s connector strip the wire leaving from 7 to 10 mm exposed and insert it into the proper terminal while pushing the fingertip down to open...

Страница 11: ...performed on a falling edge applied to the pin with respect to COMMON Low level The function remains activated while the corresponding pin is held at a low level with respect to COMMON Default factory...

Страница 12: ...scription Activation 13 ANA GROSS Makes the analog output follow the gross value measured value tare Fixed level 14 ZERO ANA Puts the analog output to the zero state 0 V for 0 10 V 4 mA for 4 20 mA Fi...

Страница 13: ...med with the MENU 3B the first digit to 1 or 2 and fourth digit to 0 During the HOLD s time on display is possible to show the stored values net gross valley peak or peak peak via keyboard or the spec...

Страница 14: ...alue in memory and goes to next step ESC Cancel programming and goes to run mode Reset peak and valley values Restart the measuring and output operation Restart the setpoints operation if 26 was OFF 1...

Страница 15: ...LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2...

Страница 16: ...djust 5V Exc 10V DC J5 J1 P2 fine adjust 10V Default setting Exc 10V P2 adjust 10V P1 adjust 5V WARNING J3 J4 J5 connect only one P1 Connection type potentiometer with Input impedance 10M POT Excitati...

Страница 17: ...is shipped from the factory with a safety code of 0000 Once introduced this code you are asked to select whether to change it or to enter directly in the parameter lockout list If you decide to chang...

Страница 18: ...PROG TE ACH PR G R UN B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN B A TAR E HO LD LIMIT MAX MIN DAT...

Страница 19: ...MIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN 19 1 Scale configuration In this menu will be intoduced the necessary para...

Страница 20: ...INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN 20 3 Display 2 value B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3...

Страница 21: ...span are 9999 and 9999 points a To validate data and advance to the next point press ENTER or b To save the programmed data in the memory and return to the run mode the meter is scaled by three point...

Страница 22: ...E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN 2 INTRODUCTION TO THE DIGIMASTER This manual does not constitute a formal agreement All information given in th...

Страница 23: ...ond the indication P 30 appears in the display 3 23 23 2 Input 30 value B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG T...

Страница 24: ...lter The figure 24 2 shows the initially selected level for the filter P any number between 0 and 9 with the FLT LED activated Press repeatedly the key to change the digit until desired value appears...

Страница 25: ...mA or lineal potentiometer for measure distance length etc The instrument supply different excitation voltages for sensors 24V 30mA 2 2V 30mA 5 or 10V 120mA allowing supply up to 4 Load Cells This dif...

Страница 26: ...values 26 This menu allows selection among 4 levels of display rounding When resolution is not critical a rounding increment higher than 1 may help to stabilize the display MENU 2AB ROUND FILTER Progr...

Страница 27: ...Indicates exit from de program mode with data memory storage 27 This menu allows enabling and disabling the tare function and its reset MENU 2 TARE The initially programmed option appears on the displ...

Страница 28: ...value different from zero is contained in the tare memory To reset the tare memory press and hold down the RESET key then press the TARE key Release first TARE then RESET To take a tare or reset it ba...

Страница 29: ...d VALEY distance value detected since the last reset or tare B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G...

Страница 30: ...m wiring connections remove the terminal block from the meter s connector strip the wire leaving from 7 to 10 mm exposed and insert it into the proper terminal while pushing the fingertip down to open...

Страница 31: ...performed on a falling edge applied to the pin with respect to COMMON Low level The function remains activated while the corresponding pin is held at a low level with respect to COMMON Default factory...

Страница 32: ...scription Activation 13 ANA GROSS Makes the analog output follow the gross value measured value tare Fixed level 14 ZERO ANA Puts the analog output to the zero state 0 V for 0 10 V 4 mA for 4 20 mA Fi...

Страница 33: ...med with the MENU 3B the first digit to 1 or 2 and fourth digit to 0 During the HOLD s time on display is possible to show the stored values net gross valley peak or peak peak via keyboard or the spec...

Страница 34: ...alue in memory and goes to next step ESC Cancel programming and goes to run mode Reset peak and valley values Restart the measuring and output operation Restart the setpoints operation if 26 was OFF 1...

Страница 35: ...LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2...

Страница 36: ...djust 5V Exc 10V DC J5 J1 P2 fine adjust 10V Default setting Exc 10V P2 adjust 10V P1 adjust 5V WARNING J3 J4 J5 connect only one P1 Connection type potentiometer with Input impedance 10M POT Excitati...

Страница 37: ...is shipped from the factory with a safety code of 0000 Once introduced this code you are asked to select whether to change it or to enter directly in the parameter lockout list If you decide to chang...

Страница 38: ...PROG TE ACH PR G R UN B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN B A TAR E HO LD LIMIT MAX MIN DAT...

Страница 39: ...MIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN 19 1 Scale configuration In this menu will be intoduced the necessary para...

Страница 40: ...INP2 FLT STO R E DSP1 INP1 2 1 4 3 TAR E R E SE T LIMIT MAX MIN E NTE R E SC DATA PROG TE ACH PR G R UN 20 3 Display 2 value B A TAR E HO LD LIMIT MAX MIN DATA DSP2 INP2 FLT STO R E DSP1 INP1 2 1 4 3...

Страница 41: ...a 30V DC F2A 250V 5 TECHNICAL SPECIFICATONS 41 Process input Input voltage pin2 versus pin 3 0 5 0 10 V Input impedance 1M Input voltage input pin1 versus pin 3 0 1V Input impedance 100M Input current...

Страница 42: ...94 V 0 Reaction time Peak value capture Input signal minimal time 2 1ms Reaction time Hold Display Max 10ms Hold Analoge Max 10ms Hold RS Max 10ms Logical Input All Max 10ms 42 Maximum and minimum inp...

Страница 43: ...he case and slide them over the guide tracks until they touch the panel at the rear side Press slightly to fasten the bezel to the panel and secure the clips To take the instrument out of the panel pu...

Страница 44: ...the warranty period please contact the distributor from whom you purchased the product This warranty does not apply to defects resulting from action of the buyer such as mishandling or improper interf...

Страница 45: ...ITY Applicable Standars EN50081 1 Generic emission EN55022 CISPR22 Class B Applicable Standars EN50082 1 Generic immunity IEC1000 4 2 Level 3 Criteria B Air Discharge 8kV Contact Discharge 6kV IEC1000...

Страница 46: ...47 OUTPUT OPTIONS RELAIS OPTOS ANNEX Valid only for 9181 See Vxxx1 Vxxx2 Vxxx3 Vxxx4 manual 53 ANALOG OUTPUT ANNEX Valid only for 9181 See Vx1xx manual...

Страница 47: ...ion in 9181 47 Menu 3B 48 Direct programming of the setpoint value 49 Output reaction time 50 TECHNICAL SPECIFICATIONS Reaction time 5ms with Filter off Cut Frequence 10Hz with Filter off Conversions...

Страница 48: ...ammed in the MENU 3B the first digit to 1 or 2 and the fourth digit to 0 When there is positive overflow OVFLO or negative overflow OVFLO The output goes to repose mode depending on the programming of...

Страница 49: ...to peak value Digit 5 0 Normaly open 1 Normaly closed NEW FUNCTIONS burster ISO FUNCTION Answer type I 0I Transmits logical input status Y 0Y Transmits Peak to peak value Returns value n 0n LATCH rela...

Страница 50: ...ng mode PROG and by pressing LIMIT key access to the setpoint 1 value The setpoints value appears at each press of ENTER key The led indicating the setpoint number turns on and the value in the main d...

Страница 51: ...100 Output reaction time 4OP 4OPP 2 1ms OPTO output Input 90 0 Setpoint 52 This time is with the 4OP or 4OPP option without filter and with the fast option in the setpoint programming see page 50...

Страница 52: ...47 OUTPUT OPTIONS RELAIS OPTOS ANNEX Valid only for 9181 See Vxxx1 Vxxx2 Vxxx3 Vxxx4 manual 53 ANALOG OUTPUT ANNEX Valid only for 9181 See Vx1xx manual...

Страница 53: ...ion in 9181 47 Menu 3B 48 Direct programming of the setpoint value 49 Output reaction time 50 TECHNICAL SPECIFICATIONS Reaction time 5ms with Filter off Cut Frequence 10Hz with Filter off Conversions...

Страница 54: ...ammed in the MENU 3B the first digit to 1 or 2 and the fourth digit to 0 When there is positive overflow OVFLO or negative overflow OVFLO The output goes to repose mode depending on the programming of...

Страница 55: ...to peak value Digit 5 0 Normaly open 1 Normaly closed NEW FUNCTIONS burster ISO FUNCTION Answer type I 0I Transmits logical input status Y 0Y Transmits Peak to peak value Returns value n 0n LATCH rela...

Отзывы: